Performance evaluation of the Sysmex haematology XN modular system
- PMID: 22851510
- DOI: 10.1136/jclinpath-2012-200930
Performance evaluation of the Sysmex haematology XN modular system
Abstract
Background: The Sysmex XN haematology instrument performs automatic reflex testing, depending on sample results. A nucleated red blood cell (NRBC) count is provided on all samples. The instrument has a smaller footprint (34%) than previous Sysmex XE analysers.
Methods: An evaluation comparing all results to the Sysmex XE-2100 and manual microscopic differential and morphology (n=390) was performed followed by a workflow study of 1000 samples to compare speed of operation and number of blood films reviews required from both systems.
Results: The new features on the instrument are: (1) white cell and NRBC channel, all samples include the NRBC count; (2) white cell precursor channel: false positive flags for blasts, abnormal lymphocytes and atypical lymphocytes are reduced significantly without a statistical increase of false negatives; (3) low white cell count mode: suggested setting of <0.5×10(9)/l. An extended count is more precise and provides an accurate differential. Fluorescent platelet count is performed in a dedicated channel. If the red cell or platelet size histograms are abnormal or if the platelet count is low, then a fluorescent platelet count is automatically performed. Good correlation with the XE-2100 and manual differential was found and the improved results compared to the reference flow cytometric analysis for platelet counts, especially below 30×10(9)/l (XE-2100, R(2)=0.500; XN, R(2)=0.875).
Conclusion: The XN showed reduced sample turnaround time of 10% and reduced number of blood films for examination, 49% less than the XE-2100 without loss of sensitivity with more precise and accurate results on low cell counts.
Similar articles
-
The value of the white precursor cell channel (WPC) on the Sysmex XN-1000 analyser in a specialist paediatric hospital.J Clin Pathol. 2015 Feb;68(2):161-5. doi: 10.1136/jclinpath-2014-202640. Epub 2014 Nov 25. J Clin Pathol. 2015. PMID: 25425659
-
Performance evaluation of the new hematology analyzer Sysmex XN-series.Int J Lab Hematol. 2015 Apr;37(2):155-64. doi: 10.1111/ijlh.12254. Epub 2014 May 12. Int J Lab Hematol. 2015. PMID: 24815300
-
Improved flagging rates on the Sysmex XE-5000 compared with the XE-2100 reduce the number of manual film reviews and increase laboratory productivity.Am J Clin Pathol. 2011 Aug;136(2):309-16. doi: 10.1309/AJCPDLR4KGKAFW4W. Am J Clin Pathol. 2011. PMID: 21757605
-
[Automated hematology analysers and spurious counts Part 2. Leukocyte count and differential].Ann Biol Clin (Paris). 2012 Mar-Apr;70(2):141-54. doi: 10.1684/abc.2012.0665. Ann Biol Clin (Paris). 2012. PMID: 22484525 Review. French.
-
Identification and quantification of high fluorescence-stained lymphocytes as antibody synthesizing/secreting cells using the automated routine hematology analyzer XE-2100.Cytometry B Clin Cytom. 2007 May;72(3):157-66. doi: 10.1002/cyto.b.20150. Cytometry B Clin Cytom. 2007. PMID: 17266152 Review.
Cited by
-
Establishment of improved review criteria for hematology analyzers in cancer hospitals.J Clin Lab Anal. 2021 Feb;35(2):e23638. doi: 10.1002/jcla.23638. Epub 2020 Dec 14. J Clin Lab Anal. 2021. PMID: 33314392 Free PMC article.
-
Comparison of four methods to measure haemoglobin concentrations in whole blood donors (COMPARE): A diagnostic accuracy study.Transfus Med. 2021 Apr;31(2):94-103. doi: 10.1111/tme.12750. Epub 2020 Dec 20. Transfus Med. 2021. PMID: 33341984 Free PMC article. Clinical Trial.
-
Sample stability for complete blood cell count using the Sysmex XN haematological analyser.Blood Transfus. 2015 Oct;13(4):576-82. doi: 10.2450/2015.0007-15. Epub 2015 Apr 27. Blood Transfus. 2015. PMID: 26057491 Free PMC article.
-
The flagging features of the Sysmex XN-10 analyser for detecting platelet clumps and the impacts of platelet clumps on complete blood count parameters.Clin Chem Lab Med. 2022 Feb 23;60(5):748-755. doi: 10.1515/cclm-2021-1226. Print 2022 Apr 26. Clin Chem Lab Med. 2022. PMID: 35212492
-
Evaluation of criteria of manual blood smear review following automated complete blood counts in a large university hospital.Rev Bras Hematol Hemoter. 2017 Oct-Dec;39(4):306-317. doi: 10.1016/j.bjhh.2017.06.007. Epub 2017 Jul 31. Rev Bras Hematol Hemoter. 2017. PMID: 29150102 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical