MEG versus EEG localization test using implanted sources in the human brain

Affiliations
Comparative Study

MEG versus EEG localization test using implanted sources in the human brain

D Cohen et al. Ann Neurol. 1990 Dec.

Abstract

It is believed that the magnetoencephalogram (MEG) localizes an electrical source in the brain to within several millimeters and is therefore more accurate than electroencephalogram (EEG) localization, reported as 20 mm. To test this belief, the localization accuracy of the MEG and EEG were directly compared. The signal source was a dipole at a known location in the brain; this was made by passing a weak current pulse simulating a neural signal through depth electrodes already implanted in patients for seizure monitoring. First, MEGs and EEGs from this dipole were measured at 16 places on the head. Then, computations were performed on the MEG and EEG data separately to determine the apparent MEG and EEG source locations. Finally, these were compared with the actual source location to determine the MEG and EEG localization errors. Measurements were made of four dipoles in each of three patients. After MEGs with weak signals were discounted, the MEG average error of localization was found to be 8 mm, which was worse than expected. The average EEG error was 10 mm, which was better than expected. These results suggest that the MEG offers no significant advantage over the EEG in localizing a focal source. However, this does not diminish other uses of the MEG.

PubMed Disclaimer

Comment in

  • MEG versus EEG localization test.
    [No authors listed] [No authors listed] Ann Neurol. 1991 Aug;30(2):222-4. doi: 10.1002/ana.410300220. Ann Neurol. 1991. PMID: 1897915 No abstract available.