Comparison of the efficacy of a custom-made pulse oximeter probe with digital electric pulp tester, cold spray, and rubber cup for assessing pulp vitality
- PMID: 22892732
- DOI: 10.1016/j.joen.2012.06.012
Comparison of the efficacy of a custom-made pulse oximeter probe with digital electric pulp tester, cold spray, and rubber cup for assessing pulp vitality
Abstract
Introduction: The ideal technique for the evaluation of pulp vitality should be noninvasive, painless, objective, reliable, and reproducible. To achieve this, the most routine tests are sensitivity tests. However, a major shortcoming with these tests is that they indirectly indicate pulp vitality by measuring a neural response. Pulse oximetry is a well-established oxygen saturation monitoring technique broadly used in medicine. However, its efficacy as the pulp vitality test should be evaluated. The aim of this study was to design and build a custom-made pulse oximeter dental probe and to evaluate its efficacy in comparison with electric pulp tester, cold spray, and a rubber cup in pulp vitality testing.
Methods: Twenty-four single-canal mandibular premolars needing endodontic treatment were selected. The patients did not have systemic disease and did not consume drugs. Also, they had no clinically relevant signs of necrosis. The selected teeth were pulpally tested with 4 kinds of tests including pulse oximetry, the electric test, cold spray, and the rubber cup. After endodontic treatment of these teeth, which revealed the actual status of the pulp, the results were analyzed by the kappa test to show the efficacy of these tests.
Results: When comparing electric, cold, heat, and pulse oximeter tests with the gold standard, the kappa agreement coefficient was 18%, 18%, 14%, and 91%, respectively. The sensitivity of pulse oximetry, a rubber cup, electric test, and cold spray was 0.93, 0.60, 0.60, and 0.53, respectively. The specificity of these tests was 1.00, 0.55, 0.22, and 0.66, respectively.
Conclusions: Pulp testing by using pulse oximetry is more reliable than the electric test, rubber cup, and cold spray. The custom-made pulse oximeter dental probe is an effective and objective method for pulp vitality assessment.
Copyright © 2012 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Evaluation of efficacy of a new custom-made pulse oximeter dental probe in comparison with the electrical and thermal tests for assessing pulp vitality.J Endod. 2007 Apr;33(4):411-4. doi: 10.1016/j.joen.2006.12.003. Epub 2007 Feb 23. J Endod. 2007. PMID: 17368329 Clinical Trial.
-
Comparison of electrical, thermal, and pulse oximetry methods for assessing pulp vitality in recently traumatized teeth.J Endod. 2007 May;33(5):531-5. doi: 10.1016/j.joen.2007.01.014. Epub 2007 Mar 12. J Endod. 2007. PMID: 17437866
-
Comparison of the reliability of laser Doppler flowmetry, pulse oximetry and electric pulp tester in assessing the pulp vitality of human teeth.J Oral Rehabil. 2011 May;38(5):340-7. doi: 10.1111/j.1365-2842.2010.02160.x. Epub 2010 Sep 26. J Oral Rehabil. 2011. PMID: 20868433
-
Dental pulp neurophysiology: part 2. Current diagnostic tests to assess pulp vitality.J Can Dent Assoc. 2009 Mar;75(2):139-43. J Can Dent Assoc. 2009. PMID: 19267966 Review.
-
Pulse oximetry: review of a potential aid in endodontic diagnosis.J Endod. 2009 Mar;35(3):329-33. doi: 10.1016/j.joen.2008.12.006. J Endod. 2009. PMID: 19249589 Review.
Cited by
-
The Validity of Pulp Tests on Crowned Teeth: A Clinical Study.Eur Endod J. 2021 Apr 22;6(2):151-4. doi: 10.14744/eej.2020.94840. Online ahead of print. Eur Endod J. 2021. PMID: 34047294 Free PMC article.
-
Comparison of the vitality test with sensitivity tests in mature and immature teeth: clinical trial.BMC Oral Health. 2024 May 27;24(1):613. doi: 10.1186/s12903-024-04317-3. BMC Oral Health. 2024. PMID: 38802767 Free PMC article. Clinical Trial.
-
Diagnostic accuracy of three placement sites for the cold test in subjects amongst different age groups.BMC Oral Health. 2019 Aug 19;19(1):189. doi: 10.1186/s12903-019-0878-2. BMC Oral Health. 2019. PMID: 31426774 Free PMC article.
-
Pulse oximetry as a dental pulp test: A scoping review to identify barriers hindering the use of oximeters in clinical practice.Saudi Dent J. 2024 Feb;36(2):262-269. doi: 10.1016/j.sdentj.2023.11.006. Epub 2023 Nov 7. Saudi Dent J. 2024. PMID: 38419999 Free PMC article.
-
Diagnostic Accuracy of Pulp Vitality Tests and Pulp Sensibility Tests for Assessing Pulpal Health in Permanent Teeth: A Systematic Review and Meta-Analysis.Int J Environ Res Public Health. 2022 Aug 4;19(15):9599. doi: 10.3390/ijerph19159599. Int J Environ Res Public Health. 2022. PMID: 35954958 Free PMC article.