Purpose: To assess the optimal b-values range for perfusion-insensitive apparent diffusion coefficient (ADC) imaging of abdominal organs using short-duration DW-MRI acquisitions with currently available ADC estimation methods.
Methods: DW-MRI data of 15 subjects were acquired with eight b-values in the range of 5-800 s∕mm(2). The reference-standard, a perfusion insensitive, ADC value (ADC(IVIM)), was computed using an intravoxel incoherent motion (IVIM) model with all acquired diffusion-weighted images. Simulated DW-MRI data was generated using an IVIM model with b-values in the range of 0-1200 s∕mm(2). Monoexponential ADC estimates were calculated using: (1) Two-point estimator (ADC(2)); (2) least squares three-point (ADC(3)) estimator and; (3) Rician noise model estimator (ADC(R)). The authors found the optimal b-values for perfusion-insensitive ADC calculations by minimizing the relative root mean square error (RRMS) between the ADC(IVIM) and the monoexponential ADC values for each estimation method and organ.
Results: Low b-value = 300 s∕mm(2) and high b-value = 1200 s∕mm(2) minimized the RRMS between the estimated ADC and the reference-standard ADC(IVIM) to less than 5% using the ADC(3) estimator. By considering only the in vivo DW-MRI data, the combination of low b-value = 270 s∕mm(2) and high b-value of 800 s∕mm(2) minimized the RRMS between the estimated ADC and the reference-standard ADC(IVIM) to <7% using the ADC(3) estimator. For all estimators, the RRMS between the estimated ADC and the reference standard ADC correlated strongly with the perfusion-fraction parameter of the IVIM model (r = [0.78-0.83], p ≤ 0.003).
Conclusions: The perfusion compartment in DW-MRI signal decay correlates strongly with the RRMS in ADC estimates from short-duration DW-MRI. The impact of the perfusion compartment on ADC estimations depends, however, on the choice of b-values and estimation method utilized. Likewise, perfusion-related errors can be reduced to <7% by carefully selecting the b-values used for ADC calculations and method of estimation.
Figures
References
-
- Stejskal E. O. and Tanner J. E., “Spin diffusion measurements: spin-echo in the presence of a time dependent field gradient,” J. Chem. Phys. 42, 288–292 (1965). 10.1063/1.1695690 - DOI
-
- Gumustas S., Inan N., Sarisoy H. T., Anik Y., Arslan A., Ciftci E., Akansel G., and Demirci A., “Malignant versus benign mediastinal lesions: quantitative assessment with diffusion weighted MR imaging,” Eur Radiol. 21, 2255–2260 (2011). - PubMed
-
- Vargas H. A., Akin O., Franiel T., Mazaheri Y., Zheng J., Moskowitz C., Udo K., Eastham J., and Hricak H., “Diffusion-weighted endorectal MR imaging at 3 T for prostate cancer: Tumor detection and assessment of aggressiveness,” Radiology 259, 775–784 (2011). 10.1148/radiol.11102066 - DOI - PMC - PubMed
-
- Dudeck O., Zeile M., Pink D., Pech M., Tunn P.-U., Reichardt P., Ludwig W.-D., and Hamm B., “Diffusion-weighted magnetic resonance imaging allows monitoring of anticancer treatment effects in patients with soft-tissue sarcomas,” J. Magn. Reson. Imaging 27, 1109–1113 (2008). 10.1002/jmri.21358 - DOI - PubMed
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources