The effects of bluephase LED light on fibroblasts
- PMID: 22904660
- PMCID: PMC3420839
The effects of bluephase LED light on fibroblasts
Abstract
Objective: The aim of this study was to evaluate the effect of Bluephase light emitting diode (LED) light on cell viability, colony-forming ability and proliferation in V79 cell culture and to determine how much the temperature of the nutrient medium rose.
Methods: The investigation included a low (L), soft start (S) and high (H) illumination mode for 20, 40 and 80 seconds. The viability was determined by the trypan blue exclusion test, colony-forming ability by counting colonies 7 days after exposure and cell proliferation by the cell counts on 5 post-exposure days. The temperature change during illumination was recorded (0.1°C sensitivity).
Results: In each experimental condition, 90-95% of the cells were viable, which was in the same range as the controls. Colony-forming ability was not found to be significantly lower (P<.05). A significant decrease in proliferation was recorded on the 4th post-exposure day with S and H irrespective of time, on the 3(rd) day with S for 80 s and H for 40 and 80 s, and with S and H for 80 s on the 2(nd) day (P<.05).The temperature rise was significant with S (P<.05) and H (P<.05), irrespective of exposure duration.
Conclusion: Dependent on total energy density, LED blue light affects the mitotic activity of cells in its path to a certain extent. Altered mitotic activity was not noted with illumination at the low power mode (intensity of 421.7 ±1.1 mW/cm(2)). The greatest temperature rise was 8.3 °C and occurred at the highest intensity and exposure duration.
Keywords: Bluephase effects; LED light; fibroblasts.
Similar articles
-
Light emitting diode-generated blue light modulates fibrosis characteristics: fibroblast proliferation, migration speed, and reactive oxygen species generation.Lasers Surg Med. 2015 Feb;47(2):210-5. doi: 10.1002/lsm.22293. Epub 2015 Feb 5. Lasers Surg Med. 2015. PMID: 25655579 Free PMC article.
-
Dose- and mode-dependent effect of halogen dental curing blue light on the V79-cell line.Coll Antropol. 2008 Sep;32(3):913-8. Coll Antropol. 2008. PMID: 18982770
-
Shrinkage evaluation of composite polymerized with LED units using laser interferometry.Quintessence Int. 2010 May;41(5):417-25. Quintessence Int. 2010. PMID: 20376378
-
Comparison of composite curing parameters: effects of light source and curing mode on conversion, temperature rise and polymerization shrinkage.Oper Dent. 2006 Mar-Apr;31(2):219-26. doi: 10.2341/05-15. Oper Dent. 2006. PMID: 16827025
-
Parameters influencing increase in pulp chamber temperature with light-curing devices: curing lights and pulpal flow rates.Oper Dent. 2010 May-Jun;35(3):353-61. doi: 10.2341/09-234-L. Oper Dent. 2010. PMID: 20533637
Cited by
-
Comparison of temperature rise in pulp chamber during polymerization of materials used for direct fabrication of provisional restorations: An in-vitro study.Eur J Dent. 2015 Apr-Jun;9(2):194-200. doi: 10.4103/1305-7456.156807. Eur J Dent. 2015. PMID: 26038649 Free PMC article.
-
Light emitting diode-generated blue light modulates fibrosis characteristics: fibroblast proliferation, migration speed, and reactive oxygen species generation.Lasers Surg Med. 2015 Feb;47(2):210-5. doi: 10.1002/lsm.22293. Epub 2015 Feb 5. Lasers Surg Med. 2015. PMID: 25655579 Free PMC article.
References
-
- Vargas MA, Cobb DS, Schmit JL. Polymerization of composite resins: argon laser vs. conventional light. Oper Dent. 1998;23:87–93. - PubMed
-
- Tarle Z, Knežević A, Demoli N, Meniga A, Šutalo J, Unterbrink G, Ristić M, Pichler G. Comparison of Composite curing parameters: effects of light source and curing mode on conversion, temperature rise and polimerization shrinkage. Oper Dent. 2006;31(2):219–226. - PubMed
-
- Jandt KD, Mills RW, Blackwell GB, Ashworth S. Depth of cure and compressive strength of dental composites cured with blue light emitting diodes (LEDs) Dent Mater. 2000;16:41–47. - PubMed
-
- Krämer N, Lohbauer U, García-Godoy F, Frankenberger R. Light curing of resin-based composites in the LED era. Am J Dent. 2008;21:135–142. - PubMed
-
- Bruzell Roll EM, Jacobsen N, Hensten-Pettersen A. Health hazards associated with curing light in the dental clinic. Clin Oral Invest. 2004;8:113–117. - PubMed