Transcatheter closure of multiple interatrial communications
- PMID: 22907918
- DOI: 10.1002/ccd.24329
Transcatheter closure of multiple interatrial communications
Abstract
Objectives: We sought to examine acute and midterm results of closure of multiple interatrial communications with staged device deployment and to review the relevant literature.
Background: Information about percutaneous methods of closure for multiple defects is limited.
Methods: We treated 148 patients with multiple defects. Of these, 88 had a relevant left to right shunt ("LRS"), 52 had a presumed paradoxical embolism ("PPE"), five had both (LRS and PPE), and one patient, respectively, had migraine, decompression sickness, and a right to left shunt. After implantation of the first device, closure of additional septal defects was attempted only if indicated clinically.
Results: Ninety-four patients received a single device and 53 more than one. In four patients, surgical defect closure followed. At the end of follow-up (FU; mean 4.5 ± 3.4 years), complete closure of all defects occurred in 67.6% (62.1% for LRS, 76.5% for PPE). Clinical success (small or trivial residual shunt) was achieved in 86.9% (83.9% for LRS, 90.2% for PPE). Complications included pericardial effusions in 2.7%, recurrent thromboembolic events in 4.8%, and new onset of atrial fibrillation in 10.1%. In a significant number of patients with multiple defects, after single device implantation, the likelihood of complete closure increased with FU time (26% complete closure at 1 month vs. 78% at 24 months).
Conclusion: Percutaneous closure of multiple interatrial communications is feasible and safe. Importantly, many residual defects close without further intervention at FU. Therefore, staged device delivery is an alternative to simultaneous device implantation, possibly requiring fewer and smaller second devices.
Copyright © 2012 Wiley Periodicals, Inc.
Similar articles
-
Five-year follow-up of transcatheter intracardiac echocardiography-assisted closure of interatrial shunts.Cardiovasc Revasc Med. 2011 Nov-Dec;12(6):355-61. doi: 10.1016/j.carrev.2011.04.003. Epub 2011 Jun 28. Cardiovasc Revasc Med. 2011. PMID: 21715232
-
Impact of occluder device type on success of percutaneous closure of atrial septal defects--a medium-term follow-up study.J Interv Cardiol. 2009 Dec;22(6):503-10. doi: 10.1111/j.1540-8183.2009.00507.x. Epub 2009 Oct 11. J Interv Cardiol. 2009. PMID: 19821880
-
Percutaneous Intervention to Treat Platypnea-Orthodeoxia Syndrome: The Toronto Experience.JACC Cardiovasc Interv. 2016 Sep 26;9(18):1928-38. doi: 10.1016/j.jcin.2016.07.003. JACC Cardiovasc Interv. 2016. PMID: 27659570
-
AMPLATZER Septal Occluder failure resulting in paradoxical cerebral embolism.Tex Heart Inst J. 2012;39(5):647-52. Tex Heart Inst J. 2012. PMID: 23109759 Free PMC article. Review.
-
Short and long term complications of device closure of atrial septal defect and patent foramen ovale: meta-analysis of 28,142 patients from 203 studies.Catheter Cardiovasc Interv. 2013 Dec 1;82(7):1123-38. doi: 10.1002/ccd.24875. Epub 2013 Aug 31. Catheter Cardiovasc Interv. 2013. PMID: 23412921 Review.
Cited by
-
Comparison of Transcatheter Atrial Septal Defect Closure in Children, Adolescents and adults: Differences, Challenges and Short-, Mid- and Long-Term Results.Korean Circ J. 2016 Nov;46(6):851-861. doi: 10.4070/kcj.2016.46.6.851. Epub 2016 Oct 20. Korean Circ J. 2016. PMID: 27826346 Free PMC article.
-
Transcatheter closure of multiple secundum atrial septal defects using multiple occluder devices: A comparative experience between pediatric and adult patients.Ann Pediatr Cardiol. 2022 Mar-Apr;15(2):128-137. doi: 10.4103/apc.apc_269_20. Epub 2022 Aug 19. Ann Pediatr Cardiol. 2022. PMID: 36246750 Free PMC article.
-
New solution for old challenge: Device closure of non-centrally positioned atrial septal defect.J Cardiovasc Thorac Res. 2020;12(3):231-233. doi: 10.34172/jcvtr.2020.40. Epub 2020 Aug 4. J Cardiovasc Thorac Res. 2020. PMID: 33123331 Free PMC article.
-
Individualized Experience With Percutaneous Transcatheter Closure of Multiple Atrial Septal Defects: A Single-Center Study.Front Cardiovasc Med. 2021 Feb 15;8:628322. doi: 10.3389/fcvm.2021.628322. eCollection 2021. Front Cardiovasc Med. 2021. PMID: 33659280 Free PMC article.
-
Fenestrated atrial septal defect percutaneously occluded by a single device: procedural and financial considerations.Cardiol Ther. 2013 Jun;2(1):97-102. doi: 10.1007/s40119-012-0009-5. Epub 2012 Nov 27. Cardiol Ther. 2013. PMID: 25135292 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources