. 2012 Nov;28(11):1126-33.
doi: 10.1016/j.dental.2012.07.160. Epub 2012 Aug 21.

Influences of implant neck design and implant-abutment joint type on peri-implant bone stress and abutment micromovement: three-dimensional finite element analysis

Affiliations

Influences of implant neck design and implant-abutment joint type on peri-implant bone stress and abutment micromovement: three-dimensional finite element analysis

Yasufumi Yamanishi et al. Dent Mater. 2012 Nov.

Abstract

Objectives: Occlusal overloading is one of the causes of peri-implant bone resorption, and many studies on stress distribution in the peri-implant bone by three-dimensional finite element analysis (3D FEA) have been performed. However, the FEA models previously reported were simplified and far from representing what occurs in clinical situations. In this study, 3D FEA was conducted with simulation of the complex structure of dental implants, and the influences of neck design and connections with an abutment on peri-implant bone stress and abutment micromovement were investigated.

Methods: Three types of two-piece implant CAD models were designed: external joint with a conical tapered neck (EJ), internal joint with a straight neck (IJ), and conical joint with a reverse conical neck (CJ). 3D FEA was performed with the setting of a "contact" condition at the component interface, and stress distribution in the peri-implant bone and abutment micromovement were analyzed.

Results: The shear stress was concentrated on the mesiodistal side of the cortical bone for EJ. EJ had the largest amount of abutment micromovement. While the von Mises and shear stresses around the implant neck were concentrated on the labial bone for IJ, they were distributed on the mesiodistal side of the cortical bone for CJ. CJ had the least amount of abutment micromovement.

Significance: Implants with a conical joint with an abutment and reverse conical neck design may effectively control occlusal overloading on the labial bone and abutment micromovement.

PubMed Disclaimer

Similar articles

Cited by

Publication types

LinkOut - more resources