Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Randomized Controlled Trial
. 2013 Jan;15(1):71-6.
doi: 10.1093/europace/eus247. Epub 2012 Aug 26.

Lead complications, device infections, and clinical outcomes in the first year after implantation of cardiac resynchronization therapy-defibrillator and cardiac resynchronization therapy-pacemaker

Collaborators, Affiliations
Randomized Controlled Trial

Lead complications, device infections, and clinical outcomes in the first year after implantation of cardiac resynchronization therapy-defibrillator and cardiac resynchronization therapy-pacemaker

Andreas Schuchert et al. Europace. 2013 Jan.

Abstract

Aims: The decision to implant a cardiac resynchronization therapy (CRT) system with (defibrillator, CRT-D) or without (pacemaker, CRT-P) cardioverter defibrillator should weigh its benefits and risks. This study examined the (i) incidence of loss of capture and infectious complications and (ii) 1-year clinical outcomes of 402 CRT-D and CRT-P recipients enrolled in the MASCOT study.

Methods and results: The indications for CRT-D or CRT-P were posed by the implanting physicians. All (i) losses of atrial and right and left ventricular capture, (ii) system-related infections, and (iii) clinical outcomes, including hospitalizations for worsening heart failure (HF) and deaths from all causes, were recorded up to 1 year of follow-up. Cardiac resynchronization therapy-defibrillator was implanted in 228 (57%) and CRT-P in 174 (43%) patients. The incidence of loss of capture was greater in CRT-D with 21 patients (9.2%) than in CRT-P with 6 patient (3.5%) recipients (P = 0.01), while the infection rates were 1.3% (3 patients) and 1.2% (2 patients), respectively (ns). In the CRT-D group, 42 of 228 patients (18.4%) died or were hospitalized for HF, compared with 38 of 174 patients (21.8%) in the CRT-P group (ns). In the CRT-D group, 23 patients (10.1%) were hospitalized for worsening HF and 20 (8.8%) patients died, vs. 22 (12.6%) and 19 (10.9%) patients, respectively, in the CRT-P group (ns for both comparisons).

Conclusions: Cardiac resynchronization therapy-defibrillator was implanted in 57% of candidates for CRT. Within 1 year after device implant, the incidence of loss of capture at any lead was nearly three-fold greater among CRT-D than among CRT-P recipients. System-related infections were infrequent and clinical outcomes were similar in both groups.

PubMed Disclaimer

Publication types

MeSH terms