Eigenspace-based minimum variance beamformer combined with Wiener postfilter for medical ultrasound imaging
- PMID: 22929478
- DOI: 10.1016/j.ultras.2012.07.012
Eigenspace-based minimum variance beamformer combined with Wiener postfilter for medical ultrasound imaging
Abstract
In this paper, a new beamformer which combines the eigenspace-based minimum variance (ESBMV) beamformer with the Wiener postfilter is proposed for medical ultrasound imaging. The primary goal of this work is to further improve the medical ultrasound imaging quality on the basis of the ESBMV beamformer. In this method, we optimize the ESBMV weights with a Wiener postfilter. With the optimization of the Wiener postfilter, the output power of the new beamformer becomes closer to the actual signal power at the imaging point than the ESBMV beamformer. Different from the ordinary Wiener postfilter, the output signal and noise power needed in calculating the Wiener postfilter are estimated respectively by the orthogonal signal subspace and noise subspace constructed from the eigenstructure of the sample covariance matrix. We demonstrate the performance of the new beamformer when resolving point scatterers and cyst phantom using both simulated data and experimental data and compare it with the delay-and-sum (DAS), the minimum variance (MV) and the ESBMV beamformer. We use the full width at half maximum (FWHM) and the peak-side-lobe level (PSL) to quantify the performance of imaging resolution and the contrast ratio (CR) to quantify the performance of imaging contrast. The FWHM of the new beamformer is only 15%, 50% and 50% of those of the DAS, MV and ESBMV beamformer, while the PSL is 127.2dB, 115dB and 60dB lower. What is more, an improvement of 239.8%, 232.5% and 32.9% in CR using simulated data and an improvement of 814%, 1410.7% and 86.7% in CR using experimental data are achieved compared to the DAS, MV and ESBMV beamformer respectively. In addition, the effect of the sound speed error is investigated by artificially overestimating the speed used in calculating the propagation delay and the results show that the new beamformer provides better robustness against the sound speed errors. Therefore, the proposed beamformer offers a better performance than the DAS, MV and ESBMV beamformer, showing its potential in medical ultrasound imaging.
Copyright © 2012 Elsevier B.V. All rights reserved.
Similar articles
-
Subarray coherence based postfilter for eigenspace based minimum variance beamformer in ultrasound plane-wave imaging.Ultrasonics. 2016 Feb;65:23-33. doi: 10.1016/j.ultras.2015.10.026. Epub 2015 Nov 10. Ultrasonics. 2016. PMID: 26582600
-
Amplitude and phase estimator combined with the Wiener postfilter for medical ultrasound imaging.J Med Ultrason (2001). 2016 Jan;43(1):11-8. doi: 10.1007/s10396-015-0671-z. Epub 2015 Oct 5. J Med Ultrason (2001). 2016. PMID: 26703162
-
Beam-domain eigenspace-based minimum variance beamformer for medical ultrasound imaging.IEEE Trans Ultrason Ferroelectr Freq Control. 2013 Dec;60(12):2670-6. doi: 10.1109/TUFFC.2013.2866. IEEE Trans Ultrason Ferroelectr Freq Control. 2013. PMID: 24297032
-
Eigenspace-based minimum variance beamforming applied to medical ultrasound imaging.IEEE Trans Ultrason Ferroelectr Freq Control. 2010 Nov;57(11):2381-90. doi: 10.1109/TUFFC.2010.1706. IEEE Trans Ultrason Ferroelectr Freq Control. 2010. PMID: 21041127
-
Detection of Point Scatterers in Medical Ultrasound.IEEE Trans Ultrason Ferroelectr Freq Control. 2022 Feb;69(2):617-628. doi: 10.1109/TUFFC.2021.3129619. Epub 2022 Jan 27. IEEE Trans Ultrason Ferroelectr Freq Control. 2022. PMID: 34797764 Review.
Cited by
-
Improvement in Multi-Angle Plane Wave Image Quality Using Minimum Variance Beamforming with Adaptive Signal Coherence.Sensors (Basel). 2024 Jan 2;24(1):262. doi: 10.3390/s24010262. Sensors (Basel). 2024. PMID: 38203125 Free PMC article.
-
2-D Minimum Variance Based Plane Wave Compounding with Generalized Coherence Factor in Ultrafast Ultrasound Imaging.Sensors (Basel). 2018 Nov 23;18(12):4099. doi: 10.3390/s18124099. Sensors (Basel). 2018. PMID: 30477114 Free PMC article.
-
Short-lag spatial coherence imaging using minimum variance beamforming on dual apertures.Biomed Eng Online. 2019 Apr 23;18(1):48. doi: 10.1186/s12938-019-0671-0. Biomed Eng Online. 2019. PMID: 31014338 Free PMC article.
-
Eigenspace-based beamformer using oblique signal subspace projection for ultrasound plane-wave imaging.Biomed Eng Online. 2016 Nov 24;15(1):127. doi: 10.1186/s12938-016-0244-4. Biomed Eng Online. 2016. PMID: 27881172 Free PMC article.
-
Improving Ultrasound B-Mode Image Quality with Coherent Plane-Wave Compounding Using Adaptive Beamformers Based on Minimum Variance.Sensors (Basel). 2025 Feb 21;25(5):1306. doi: 10.3390/s25051306. Sensors (Basel). 2025. PMID: 40096015 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources