. 2012 Dec;65(6):513-20.
doi: 10.1016/j.jinf.2012.08.013. Epub 2012 Aug 30.

Identification and susceptibility testing of microorganism by direct inoculation from positive blood culture bottles by combining MALDI-TOF and Vitek-2 Compact is rapid and effective

Affiliations

Identification and susceptibility testing of microorganism by direct inoculation from positive blood culture bottles by combining MALDI-TOF and Vitek-2 Compact is rapid and effective

María-Pilar Romero-Gómez et al. J Infect. 2012 Dec.

Abstract

Objectives: The objective of this study was to evaluate the reliability and accuracy of the combined use of MALDI-TOF MS bacterial identification and the Vitek-2 Compact antimicrobial susceptibility testing (AST) directly from positive blood cultures.

Methods: Direct identification by MALDI-TOF MS and AST were performed in parallel to the standard methods in all positively flagged blood cultures bottles during the study period.

Results: Three hundred and twenty four monomicrobial positive blood cultures were included in the present study, with 257 Gram-negative and 67 Gram-positive isolates. MALDI-TOF MS identification directly from blood bottles reported the correct identification for Enterobacteriaceae in 97.7%, non-fermentative Gram-negative bacilli 75.0%, Staphylococcus aureus 75.8%, coagulase negative staphylococci 63.3% and enterococci 63.3%. A total 6156 isolate/antimicrobial agent combinations were tested. Enterobacteriaceae group and non-fermentative Gram-negative Bacilli showed an agreement of 96.67% and 92.30%, respectively, for the Gram-positive cocci the overall agreement found was 97.84%.

Conclusions: We conclude that direct identification by MALDI-TOF and inoculation of Vitek-2 Compact AST with positive blood culture bottles yielded very good results and decreased time between initial inoculation of blood culture media and determination of the antibiotic susceptibility for Gram-negative rods and Gram-positive cocci causing bacteremia.