. 2012 Aug;121(8):549-54.
doi: 10.1177/000348941212100809.

Ethnic disparity in skin complications following bone-anchored hearing aid implantation

Affiliations

Ethnic disparity in skin complications following bone-anchored hearing aid implantation

Daniel M Zeitler et al. Ann Otol Rhinol Laryngol. 2012 Aug.

Abstract

Objectives: Sound processor loading after implantation of a bone-anchored hearing aid is often delayed by skin-site complications. This study examined the frequency of skin-site complications in various ethnic groups and determined factors that may lead to higher rates of skin-site complications resulting in delayed processor loading.

Methods: Adult, English-speaking patients who underwent implantation of a bone-anchored hearing aid between 2007 and 2010 were reviewed. Demographic data including ethnicity, tobacco use, diabetes mellitus, immunosuppression, and long-term steroid use were determined. Major and minor skin-site complications and the time to processor loading were recorded.

Results: The mean time to processor loading was 9.5 weeks, and the mean follow-up time was 23 months. There were no cases of osseointegration failure. African American patients had a significantly higher rate of major skin-site complications (p < 0.005) and a longer time to processor loading (mean, 17.6 weeks; p < 0.05) than the other ethnic groups. There was no significant difference in minor skin complication rates. There was no correlation between diabetes mellitus, long-term immunosuppression, or tobacco use and skin-site complications.

Conclusions: Skin complications can delay processor loading following implantation of a bone-anchored hearing aid. There is a higher rate of major skin-site complications in African American patients, and these often delay processor loading. The risk of skin-site complications is not correlated with smoking, diabetes mellitus, or immunosuppression. An increased risk of skin-site complications is an important consideration for preoperative counseling.