The use and efficacy of continuous glucose monitoring in type 1 diabetes treated with insulin pump therapy: a randomised controlled trial
- PMID: 22965294
- PMCID: PMC3483098
- DOI: 10.1007/s00125-012-2708-9
The use and efficacy of continuous glucose monitoring in type 1 diabetes treated with insulin pump therapy: a randomised controlled trial
Abstract
Aims/hypothesis: The aim of this multicentre, randomised, controlled crossover study was to determine the efficacy of adding continuous glucose monitoring (CGM) to insulin pump therapy (CSII) in type 1 diabetes.
Methods: Children and adults (n = 153) on CSII with HbA(1c) 7.5-9.5% (58.5-80.3 mmol/mol) were randomised to (CGM) a Sensor On or Sensor Off arm for 6 months. After 4 months' washout, participants crossed over to the other arm for 6 months. Paediatric and adult participants were separately electronically randomised through the case report form according to a predefined randomisation sequence in eight secondary and tertiary centres. The primary outcome was the difference in HbA(1c) levels between arms after 6 months.
Results: Seventy-seven participants were randomised to the On/Off sequence and 76 to the Off/On sequence; all were included in the primary analysis. The mean difference in HbA(1c) was -0.43% (-4.74 mmol/mol) in favour of the Sensor On arm (8.04% [64.34 mmol/mol] vs 8.47% [69.08 mmol/mol]; 95% CI -0.32%, -0.55% [-3.50, -6.01 mmol/mol]; p < 0.001). Following cessation of glucose sensing, HbA(1c) reverted to baseline levels. Less time was spent with sensor glucose <3.9 mmol/l during the Sensor On arm than in the Sensor Off arm (19 vs 31 min/day; p = 0.009). The mean number of daily boluses increased in the Sensor On arm (6.8 ± 2.5 vs 5.8 ± 1.9, p < 0.0001), together with the frequency of use of the temporary basal rate (0.75 ± 1.11 vs 0.26 ± 0.47, p < 0.0001) and manual insulin suspend (0.91 ± 1.25 vs 0.70 ± 0.75, p < 0.018) functions. Four vs two events of severe hypoglycaemia occurred in the Sensor On and Sensor Off arm, respectively (p = 0.40).
Conclusions/interpretation: Continuous glucose monitoring was associated with decreased HbA(1c) levels and time spent in hypoglycaemia in individuals with type 1 diabetes using CSII. More frequent self-adjustments of insulin therapy may have contributed to these effects.
Trial registration: ClinicalTrials.gov NCT00598663.
Figures


Similar articles
-
Effect of initiating use of an insulin pump in adults with type 1 diabetes using multiple daily insulin injections and continuous glucose monitoring (DIAMOND): a multicentre, randomised controlled trial.Lancet Diabetes Endocrinol. 2017 Sep;5(9):700-708. doi: 10.1016/S2213-8587(17)30217-6. Epub 2017 Jul 12. Lancet Diabetes Endocrinol. 2017. PMID: 28711468 Clinical Trial.
-
Glycaemic impact of patient-led use of sensor-guided pump therapy in type 1 diabetes: a randomised controlled trial.Diabetologia. 2009 Jul;52(7):1250-7. doi: 10.1007/s00125-009-1365-0. Epub 2009 Apr 25. Diabetologia. 2009. PMID: 19396424 Clinical Trial.
-
Sensor-augmented pump therapy lowers HbA(1c) in suboptimally controlled Type 1 diabetes; a randomized controlled trial.Diabet Med. 2011 Oct;28(10):1158-67. doi: 10.1111/j.1464-5491.2011.03256.x. Diabet Med. 2011. PMID: 21294770 Clinical Trial.
-
Severe hypoglycaemia and glycaemic control in Type 1 diabetes: meta-analysis of multiple daily insulin injections compared with continuous subcutaneous insulin infusion.Diabet Med. 2008 Jul;25(7):765-74. doi: 10.1111/j.1464-5491.2008.02486.x. Diabet Med. 2008. PMID: 18644063 Review.
-
Meta-analysis of individual patient data in randomised trials of self monitoring of blood glucose in people with non-insulin treated type 2 diabetes.BMJ. 2012 Feb 27;344:e486. doi: 10.1136/bmj.e486. BMJ. 2012. PMID: 22371867
Cited by
-
Intermittently Scanned and Continuous Glucose Monitor Systems: A Systematic Review on Psychological Outcomes in Pediatric Patients.Front Pediatr. 2021 May 5;9:660173. doi: 10.3389/fped.2021.660173. eCollection 2021. Front Pediatr. 2021. PMID: 34026692 Free PMC article.
-
HypoDE: Research Design and Methods of a Randomized Controlled Study Evaluating the Impact of Real-Time CGM Usage on the Frequency of CGM Glucose Values <55 mg/dl in Patients With Type 1 Diabetes and Problematic Hypoglycemia Treated With Multiple Daily Injections.J Diabetes Sci Technol. 2015 May;9(3):651-62. doi: 10.1177/1932296815575999. Epub 2015 Mar 9. J Diabetes Sci Technol. 2015. PMID: 25759183 Free PMC article. Clinical Trial.
-
Routine use of continuous glucose monitoring in 10 501 people with diabetes mellitus.Diabet Med. 2015 Dec;32(12):1568-74. doi: 10.1111/dme.12825. Epub 2015 Jul 14. Diabet Med. 2015. PMID: 26042926 Free PMC article.
-
Evaluating a systematic intensive therapy using continuous glucose monitoring and intermittent scanning glucose monitoring in clinical diabetes care: a protocol for a multi-center randomized clinical trial.Front Clin Diabetes Healthc. 2023 Oct 18;4:1247616. doi: 10.3389/fcdhc.2023.1247616. eCollection 2023. Front Clin Diabetes Healthc. 2023. PMID: 37920192 Free PMC article.
-
Impact of Human Factors Testing on Medical Device Design: Validation of an Automated CGM Sensor Applicator.J Diabetes Sci Technol. 2019 Sep;13(5):949-953. doi: 10.1177/1932296819831071. Epub 2019 Feb 14. J Diabetes Sci Technol. 2019. PMID: 30762437 Free PMC article.
References
-
- Diabetes Control and Complications Trial/Epidemiology of Diabetes Interventions and Complications (DCCT/EDIC) Research Group. Nathan DM, Zinman B, et al. Modern-day clinical course of type 1 diabetes mellitus after 30 years’ duration: the diabetes control and complications trial/epidemiology of diabetes interventions and complications and Pittsburgh epidemiology of diabetes complications experience (1983-2005) Arch Intern Med. 2000;169:1307–1316. - PMC - PubMed
-
- Misso ML, Egberts KJ, Page M, O’Connor D, Shaw J (2010) Continuous subcutaneous insulin infusion (CSII) versus multiple insulin injections for type 1 diabetes mellitus. Cochrane Database Syst Rev, Issue 1, Art. No.: doi:10.1002/14651858.CD005103.pub2 - PubMed
Publication types
MeSH terms
Substances
Associated data
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous