. 2012 Nov;265(2):544-54.
doi: 10.1148/radiol.12112201. Epub 2012 Sep 10.

Radiation exposure from CT scans: how to close our knowledge gaps, monitor and safeguard exposure--proceedings and recommendations of the Radiation Dose Summit, sponsored by NIBIB, February 24-25, 2011

Affiliations

Radiation exposure from CT scans: how to close our knowledge gaps, monitor and safeguard exposure--proceedings and recommendations of the Radiation Dose Summit, sponsored by NIBIB, February 24-25, 2011

John M Boone et al. Radiology. 2012 Nov.

Abstract

This article summarizes the proceedings of a portion of the Radiation Dose Summit, which was organized by the National Institute of Biomedical Imaging and Bioengineering and held in Bethesda, Maryland, in February 2011. The current understandings of ways to optimize the benefit-risk ratio of computed tomography (CT) examinations are summarized and recommendations are made for priority areas of research to close existing gaps in our knowledge. The prospects of achieving a submillisievert effective dose CT examination routinely are assessed.

PubMed Disclaimer

Figures

Figure 1:
Figure 1:
The x-axis represents a patient size metric (effective diameter) and the y-axis represents the normalized dose coefficient for the 32 cm polymethyl methacrylate (PMMA) phantom. Note that AAPM Task Group 204 Report also contains conversions for 16 cm PMMA phantom as well. (Reprinted, with permission, from reference .)
Figure 2:
Figure 2:
Left: In perfusion or interventional CT (where there is no table movement), the peak skin dose is the relevant dose parameter for deterministic skin effects. The peak skin dose equals the peak dose from one scan times the number of scans. Right: If the peripheral CTDI100 is used as a surrogate for peak dose, the skin dose will be overestimated by up to a factor of two. (Reprinted, with permission, from reference .)

References

    1. Los Angeles Times Cedars-Sinai investigated for significant radiation overdoses of 206 patients. http://latimesblogs.latimes.com/lanow/2009/10/cedarssinai-investigated-f.... Published October 9, 2009. Accessed April 17, 2011
    1. McKinnon PJ. Ataxia-telangiectasia: an inherited disorder of ionizing-radiation sensitivity in man—progress in the elucidation of the underlying biochemical defect. Hum Genet 1987;75(3):197–208 - PubMed
    1. Hricak H, Brenner DJ, Adelstein SJ, et al. Managing radiation use in medical imaging: a multifaceted challenge. Radiology 2011;258(3):889–905 - PubMed
    1. National Research Council Health risks from exposure to low levels of ionizing radiation: BEIR VII–Phase 2. National Academy of Sciences. Washington, DC: National Academies Press, 2006 - PubMed
    1. Image Gently The Alliance for Radiation Safety in Pediatric Imaging. http://www.pedrad.org/associations/5364/ig/. Accessed April 17, 2011