Direct measurement of mammographic X-ray spectra with a digital CdTe detection system
- PMID: 22969406
- PMCID: PMC3436035
- DOI: 10.3390/s120608390
Direct measurement of mammographic X-ray spectra with a digital CdTe detection system
Abstract
In this work we present a detection system, based on a CdTe detector and an innovative digital pulse processing (DPP) system, for high-rate X-ray spectroscopy in mammography (1-30 keV). The DPP system performs a height and shape analysis of the detector pulses, sampled and digitized by a 14-bit, 100 MHz ADC. We show the results of the characterization of the detection system both at low and high photon counting rates by using monoenergetic X-ray sources and a nonclinical X-ray tube. The detection system exhibits excellent performance up to 830 kcps with an energy resolution of 4.5% FWHM at 22.1 keV. Direct measurements of clinical molybdenum X-ray spectra were carried out by using a pinhole collimator and a custom alignment device. A comparison with the attenuation curves and the half value layer values, obtained from the measured and simulated spectra, from an ionization chamber and from a solid state dosimeter, also shows the accuracy of the measurements. These results make the proposed detection system a very attractive tool for both laboratory research, calibration of dosimeters and advanced quality controls in mammography.
Keywords: CdTe detectors; X-ray spectroscopy; digital pulse processing; high photon counting rate; mammography.
Figures







References
-
- Sanborg M., Dance D.R., Alm Carlsson G., Persliden J. A Monte Carlo for the simulation of image quality and absorbed dose in diagnostic radiology. Comput. Methods Prog. Biomed. 1994;31:167–180. - PubMed
-
- Boone J.M., Fewell T.R., Jennings R.J. Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography. Med. Phys. 1997;24:1863–1874. - PubMed
-
- Sidky E.Y., Yu L., Pan X., Zou Y., Vannier M. A robust method of X-ray source spectrum estimation from transmission measurements: Demonstrated on computer simulated, scatter-free transmission data. J. Appl. Phys. 2005;97:124701.
-
- Silva M.C., Herdade S.B., Lammoglia P., Costa P.R., Terini R.A. Determination of the voltage applied to X-ray tubes from the bremsstrahlung spectrum obtained with a silicon PIN photodiode. Med. Phys. 2000;27:2617–2623. - PubMed
-
- Assiamah M., Nam T.L., Keddy R.J. Comparison of mammography radiation dose values obtained from direct incident air kerma measurements with values from measured X-ray spectral data. Appl. Radiat. Isot. 2005;52:551–560. - PubMed
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical