Corneal thickness after overnight wear of an intraocular pressure fluctuation contact lens sensor
- PMID: 22974389
- DOI: 10.1111/j.1755-3768.2012.02495.x
Corneal thickness after overnight wear of an intraocular pressure fluctuation contact lens sensor
Abstract
Purpose: To assess the effect of overnight wear of a contact lens-based sensor (CLS) for monitoring of 24-hr intraocular pressure (IOP) fluctuations on central corneal thickness (CCT).
Materials and methods: Changes in the CCT, mid-peripheral corneal thickness and central corneal radius (CCR) during overnight CLS wear in 20 eligible patients with ocular hypertension or established glaucoma were prospectively studied using ultrasound pachymetry and topography. Corneal thickness and CCR changes were evaluated from pre-to-postsleep, with the fellow eye as control. Paired t-test or Wilcoxon signed-rank test was used as appropriate and with α = 0.05. Relationship between the IOP profile recorded by the CLS and the pre-to-postsleep corneal thickness differences was assessed using the Spearman correlation coefficient.
Results: After CLS wear, mean CCT had changed from 523 to 537 μm (p = 0.015) in the study eye and from 518 to 522 (p = 0.206) in the fellow eye (n = 15). There was no difference in CCT change between eyes (p = 0.075). There were no statistically significant changes in horizontal or vertical CCR in either eye (p > 0.05 for all). No correlation was found between the pre-to-postsleep differences in the CLS signal and the pre-to-postsleep differences in ultrasound CCT measurements (p = 0.974).
Conclusion: The continuous IOP monitoring does not appear to be affected by differences in corneal thickness that occur during overnight CLS wear, although the CLS did induce some corneal swelling. This effect was not statistically significantly different from the control eye and does not seem to influence the CLS IOP profile.
© 2012 The Authors. Acta Ophthalmologica © 2012 Acta Ophthalmologica Scandinavica Foundation.
Similar articles
-
Effect of overnight wear of the Triggerfish(®) sensor on corneal thickness measured by Visante(®) anterior segment optical coherence tomography.Acta Ophthalmol. 2014 Mar;92(2):e119-23. doi: 10.1111/aos.12241. Epub 2013 Jul 15. Acta Ophthalmol. 2014. PMID: 23848265
-
Changes in corneal biomechanical properties after 24 hours of continuous intraocular pressure monitoring using a contact lens sensor.Can J Ophthalmol. 2018 Jun;53(3):236-241. doi: 10.1016/j.jcjo.2017.10.028. Can J Ophthalmol. 2018. PMID: 29784159
-
Influence of Ocular Dimensional Change on 24-Hour Intraocular Pressure Measurement With Contact Lens Sensor.J Glaucoma. 2019 Sep;28(9):808-810. doi: 10.1097/IJG.0000000000001318. J Glaucoma. 2019. PMID: 31233460
-
Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach.Surv Ophthalmol. 2000 Mar-Apr;44(5):367-408. doi: 10.1016/s0039-6257(00)00110-7. Surv Ophthalmol. 2000. PMID: 10734239 Review.
-
Outcome, influence factor and development of CLS measurement in continuous IOP monitoring: A narrative review.Cont Lens Anterior Eye. 2021 Aug;44(4):101376. doi: 10.1016/j.clae.2020.10.006. Epub 2020 Oct 19. Cont Lens Anterior Eye. 2021. PMID: 33092960 Review.
Cited by
-
Comparison of intraocular pressure fluctuations before and after ab interno trabeculectomy in pseudoexfoliation glaucoma patients.Clin Ophthalmol. 2017 Sep 14;11:1667-1675. doi: 10.2147/OPTH.S143061. eCollection 2017. Clin Ophthalmol. 2017. PMID: 28979095 Free PMC article.
-
Transient changes in refractive error and corneal tomography after 24-h continuous monitoring of intraocular pressure patterns with a contact lens sensor.Jpn J Ophthalmol. 2020 Mar;64(2):127-133. doi: 10.1007/s10384-020-00723-6. Epub 2020 Feb 13. Jpn J Ophthalmol. 2020. PMID: 32056036
-
The road ahead to continuous 24-hour intraocular pressure monitoring in glaucoma.J Ophthalmic Vis Res. 2014 Apr;9(2):260-8. J Ophthalmic Vis Res. 2014. PMID: 25279129 Free PMC article.
-
Twenty-Four-Hour Contact Lens Sensor Monitoring of Aqueous Humor Dynamics in Surgically or Medically Treated Glaucoma Patients.J Ophthalmol. 2019 Jan 27;2019:9890831. doi: 10.1155/2019/9890831. eCollection 2019. J Ophthalmol. 2019. PMID: 30809389 Free PMC article.
-
Estimation of 24-Hour Intraocular Pressure Peak Timing and Variation Using a Contact Lens Sensor.PLoS One. 2015 Jun 15;10(6):e0129529. doi: 10.1371/journal.pone.0129529. eCollection 2015. PLoS One. 2015. PMID: 26076472 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical