Potential of dosage reduction in cone-beam-computed tomography (CBCT) for radiological diagnostics of the paranasal sinuses

Affiliations

Potential of dosage reduction in cone-beam-computed tomography (CBCT) for radiological diagnostics of the paranasal sinuses

C Güldner et al. Eur Arch Otorhinolaryngol. 2013 Mar.

Abstract

More than 10 years ago, cone-beam-computed tomography (CBCT) was introduced in ENT radiology. Until now, the focus of research was to evaluate clinical limits of this technique. The aim of this work is the evaluation of specific dosages and the identification of potential optimization in the performance of CBCT of the paranasal sinuses. Based on different tube parameters (tube current, tube voltage, and rotation angles), images of the nose and the paranasal sinuses were taken on a phantom head with the Accu-I-tomo F17 (Morita, Kyoto, Japan). The dosages applied to the lens and parotid gland were measured with OSL dosimetry. The imaging quality was evaluated by independent observers. All datasets were reviewed according to a checklist of surgically important anatomic structures. Even for lowest radiation exposure (4 mA, 76 kV, 180°, computed tomography dosage index (CTDI) = 1.8 mGy), the imaging quality was sufficient. Of course a significant reduction of the imaging quality could be seen, so a reliable mean was set for 4 mA, 84 kV, and 180° rotation angle (CTDI = 2.4 mGy). In this combination, a reduction of 92 % in lens-dose and of 77 % of dosage at the parotid gland was observed in comparison to the maximal possible adjustments (8 mA, 90 kV, 360°, CTDI = 10.9 mGy). There is potential for optimization in CBCT. Changing the rotation angle (180° instead of 360°) leads to a dose reduction of 50 %. Furthermore from clinical point of view in case of chronic rhinosinusitis a relevant reduction of dosage is possible. Therefore, it is necessary to intensify the interdisciplinary discussion about the disease specifics required quality of imaging.

PubMed Disclaimer

References

    1. Clin Otolaryngol. 2006 Jun;31(3):187-91 - PubMed
    1. Laryngol Rhinol Otol (Stuttg). 1988 Apr;67(4):142-9 - PubMed
    1. HNO. 2012 Feb;60(2):141-62 - PubMed
    1. Phys Med Biol. 2008 Oct 21;53(20):R351-79 - PubMed
    1. Z Laryngol Rhinol Otol. 1962 Nov;41:809-13 - PubMed

MeSH terms

LinkOut - more resources