Dynamic MR imaging of a minipig's knee using a high-density multi-channel receive array and a movement device
- PMID: 23014944
- DOI: 10.1007/s10334-012-0341-8
Dynamic MR imaging of a minipig's knee using a high-density multi-channel receive array and a movement device
Abstract
Object: To construct an optimised, high-density receive array and a movement device to achieve dynamic imaging of the knee in orthopedic large animal models (e.g., minipigs) at 1.5 T.
Materials and methods: A 13-channel RF receive array was constructed, and the crucial choice of the array element size (based on considerations like region of interest, geometry of the minipig's knee, achievable signal-to-noise ratio, applicability of parallel imaging, etc.) was determined using the Q factors of loops with different sizes. A special movement device was constructed to guide and produce a reproducible motion of the minipig's knee during acquisition.
Results: The constructed array was electrically characterised and the reproducibility of the cyclic motion was validated. Snapshots of dynamic in vivo images taken at a temporal resolution (308 ms) are presented. Some of the fine internal structures within the minipig's knee, like cruciate ligaments, are traced in the snapshots.
Conclusion: This study is a step towards making dynamic imaging which can give additional information about joint injuries when static MRI is not able to give sufficient information, a routine clinical application. There, the combination of a high-density receive array and a movement device will be highly helpful in the diagnosis and therapy monitoring of knee injuries in the future.