Evaluation of a fiberoptic intracranial pressure monitor
- PMID: 2303881
- DOI: 10.3171/jns.1990.72.3.0482
Evaluation of a fiberoptic intracranial pressure monitor
Abstract
The fiberoptic device is a relatively new type of intracranial pressure (ICP) monitor which appears to offer certain advantages over conventional monitoring systems, particularly its ability to measure brain parenchymal pressures. This study was undertaken to analyze the accuracy and drift characteristics of the fiberoptic device and to compare pressures in the subdural, intraparenchymal, and intraventricular compartments. The device was accurate to +/- 3 mm Hg over a 0- to 30-mm Hg range in vitro. The maximum daily drift was +/- 2.5 mm Hg, with an average daily drift of +/- 0.6 mm Hg and an average drift over a 5-day period of +/- 2.1 mm Hg. In vivo, the pressures and waveform characteristics obtained with the fiberoptic device and with a strain-gauge transducer connected to a ventriculostomy were very similar. Alterations in ICP were induced by various therapeutic and pathological manipulations, and the pressures in the three intracranial compartments were compared. Changes in ICP appeared to be reflected simultaneously and equally in all three compartments. Furthermore, changes in ICP secondary to a unilateral mass lesion were identical in both supratentorial parenchymal compartments when measured simultaneously. It is concluded that the fiberoptic device is an accurate and reliable system for ICP monitoring; the pressures recorded in the subdural, intraparenchymal, and intraventricular compartments paralleled each other in all of the physiological and pathological states tested. Although the drift associated with this device is less than that reported for previously available systems, its maximal cumulative drift over a 5-day period of +/- 6 mm Hg is significant. Since the fiberoptic device cannot be recalibrated in situ, it is suggested that the device be replaced if monitoring is to be continued for periods longer than 5 days.
Comment in
-
Fiberoptic pressure monitors.J Neurosurg. 1990 Oct;73(4):642. J Neurosurg. 1990. PMID: 2398398 No abstract available.
Similar articles
-
Experimental evaluation of the Spiegelberg intracranial pressure and intracranial compliance monitor. Technical note.J Neurosurg. 2000 Dec;93(6):1072-7. doi: 10.3171/jns.2000.93.6.1072. J Neurosurg. 2000. PMID: 11117854
-
Monitoring of brain tissue pressure with a fiberoptic device.Neurosurgery. 1992 Nov;31(5):918-21; discussion 921-2. doi: 10.1227/00006123-199211000-00014. Neurosurgery. 1992. PMID: 1436417
-
The Camino intracranial pressure device in clinical practice: reliability, handling characteristics and complications.Acta Neurochir (Wien). 1998;140(11):1113-9; discussion 1119-20. doi: 10.1007/s007010050224. Acta Neurochir (Wien). 1998. PMID: 9870055
-
The Brain Trauma Foundation. The American Association of Neurological Surgeons. The Joint Section on Neurotrauma and Critical Care. Recommendations for intracranial pressure monitoring technology.J Neurotrauma. 2000 Jun-Jul;17(6-7):497-506. doi: 10.1089/neu.2000.17.497. J Neurotrauma. 2000. PMID: 10937892 Review.
-
Intracranial pressure monitoring techniques.New Horiz. 1995 Aug;3(3):395-9. New Horiz. 1995. PMID: 7496747 Review.
Cited by
-
Comparative analysis between epidural (Gaeltec) and subdural (Camino) intracranial pressure probes.J Clin Monit. 1992 Apr;8(2):116-20. doi: 10.1007/BF01617429. J Clin Monit. 1992. PMID: 1583546
-
Intraoperative measurement of intraventricular pressure in dogs with communicating internal hydrocephalus.PLoS One. 2019 Sep 27;14(9):e0222725. doi: 10.1371/journal.pone.0222725. eCollection 2019. PLoS One. 2019. PMID: 31560704 Free PMC article.
-
Complications of brain tissue pressure monitoring with a fiberoptic device.Neurosurg Rev. 1998;21(4):254-9. doi: 10.1007/BF01105781. Neurosurg Rev. 1998. PMID: 10068186
-
Hypoxia mediated pulmonary edema: potential influence of oxidative stress, sympathetic activation and cerebral blood flow.BMC Physiol. 2015 Oct 9;15:4. doi: 10.1186/s12899-015-0018-4. BMC Physiol. 2015. PMID: 26449218 Free PMC article.
-
A comparison of extradural and intraparenchymatous intracranial pressures in head injured patients.Intensive Care Med. 1995 Oct;21(10):850-2. doi: 10.1007/BF01700971. Intensive Care Med. 1995. PMID: 8557876 Clinical Trial.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources