Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Clinical Trial
. 2013 Apr;101(4):1144-57.
doi: 10.1002/jbm.a.34392. Epub 2012 Oct 9.

Influence of a layer-by-layer-assembled multilayer of anti-CD34 antibody, vascular endothelial growth factor, and heparin on the endothelialization and anticoagulation of titanium surface

Affiliations
Clinical Trial

Influence of a layer-by-layer-assembled multilayer of anti-CD34 antibody, vascular endothelial growth factor, and heparin on the endothelialization and anticoagulation of titanium surface

Shihui Liu et al. J Biomed Mater Res A. 2013 Apr.

Abstract

The endothelialization of the metal surface of vascular stents came into focus as a new method for improving the biocompatibility of intravascular stents. This article has its focus on building a biofunctional layer on the activated titanium surface with anti-CD34 antibody, vascular endothelial growth factor (VEGF), and heparin by a layer-by-layer (LBL) self-assembly technique, to promote the endothelialization of the surface. When compared with titanium surface, the number of adhered endothelial progenitor cells (EPCs) on LBLs increased 47.1% after 5 days culture, meanwhile the proliferation rate of EPCs on LBLs during 5 days culture also exhibit significantly improvement. The results of blood compatibility evaluation clearly show that the LBLs reduce the number of adhered and activated blood platelets (adhered: Ti 83% vs. LBL <20% and activated: Ti 57% vs. LBL 19%), and the activated partial thromboplastin time of the LBL surface prolonged about 20 s in compared with platelet-poor plasma. The assembled LBL thus improves the thromboresistance of the titanium surface. The presented biofunctional multilayer of anti-CD34, VEGF, and heparin on titanium can significantly improve the blood compatibility and the endothelialization of a medical device. The biofunctional layer supports generation of a new endothelium onto titanium surface by capturing EPCs and oriented differentiation.