Esophageal pulse oximetry is more accurate and detects hypoxemia earlier than conventional pulse oximetry during general anesthesia
- PMID: 23054501
- DOI: 10.1007/s11684-012-0217-3
Esophageal pulse oximetry is more accurate and detects hypoxemia earlier than conventional pulse oximetry during general anesthesia
Abstract
The esophagus is perfused directly by prominent arteries and may provide a more consistent tissue source for pulse oximetry. The goal of this study was to evaluate the sensitivity and accuracy of an esophageal pulse oximetry probe on patients during controlled hypoxemia in comparison to measurements obtained with conventional pulse oximetry (SpulseO(2)). Forty-five ASA I-II adult patients were included in this prospective observational study. Nellcor digital oximetric probes were placed on finger tips for SpulseO(2) before anesthesia. After tracheal intubation, an esophageal probe was placed in the lower segment of the esophagus for esophageal oximetric monitoring (SoesO(2)). All patients were disconnected from the breathing circuit to establish a controlled hypoxemia, and were re-connected to the breathing circuit and ventilated with 100% oxygen immediately when SoesO(2) dropped to 90%. Matched SoesO(2) and SpulseO(2) readings were recorded when SoesO(2) measurements were at 100%, 95%, 90% and the lowest reading. The time for SoesO(2) and SpulseO(2) to drop from 100% to 95%, 90% and return to 100% was recorded. Oxygen saturation from arterial blood samples (SartO(2)) was also measured at each time point respectively. The linear correlation coefficient of the regression analysis between SartO(2) and SoesO(2) was 0.954. The mean ± 2SD of the difference was 0.3% ± 4.3% for SoesO(2) vs. SartO(2) and 6.8% ± 5.6% for SpulseO(2) vs. SartO(2) (P < 0.001). The 95% confidence interval for the absolute difference between SoesO(2) and SartO(2) was 0.3% to 0.7% and 6.2% to 7.4% between SpulseO(2) and SartO(2). The time to reach 90% saturation measured with SoesO(2) was approximately 94 seconds earlier than the SpulseO(2) (P < 0.001). In conclusion, SoesO(2) is more accurate and enables earlier detection of hypoxemia when compared to conventional pulse oximetry during hypoxemia for patients undergoing general anesthesia.
Similar articles
-
Comparison of a prototype esophageal oximetry probe with two conventional digital pulse oximetry monitors in aortocoronary bypass patients.J Clin Monit Comput. 2000;16(3):201-9. doi: 10.1023/a:1009941610320. J Clin Monit Comput. 2000. PMID: 12578104
-
[Hypoxemia after general anesthesia].Anaesthesist. 2012 Apr;61(4):299-309. doi: 10.1007/s00101-012-2000-x. Anaesthesist. 2012. PMID: 22526741 German.
-
Accuracy of calculated arterial saturation in oxygen in neonatal foals and effects of monitor, sensor, site of sensor placement, and degree of hypoxemia on the accuracy of pulse oximetry.J Vet Emerg Crit Care (San Antonio). 2014 Sep-Oct;24(5):529-35. doi: 10.1111/vec.12231. Epub 2014 Sep 11. J Vet Emerg Crit Care (San Antonio). 2014. PMID: 25212699
-
[Development of pulse oximetry].Nihon Kyobu Shikkan Gakkai Zasshi. 1994 Dec;32 Suppl:96-100. Nihon Kyobu Shikkan Gakkai Zasshi. 1994. PMID: 7602851 Review. Japanese.
-
Advanced Uses of Pulse Oximetry for Monitoring Mechanically Ventilated Patients.Anesth Analg. 2017 Jan;124(1):62-71. doi: 10.1213/ANE.0000000000001283. Anesth Analg. 2017. PMID: 27183375 Review.
Cited by
-
Evaluation of a Novel Ear Pulse Oximeter: Towards Automated Oxygen Titration in Eyeglass Frames.Sensors (Basel). 2020 Jun 10;20(11):3301. doi: 10.3390/s20113301. Sensors (Basel). 2020. PMID: 32531975 Free PMC article.
-
Esophageal oxyhemoglobin saturation as a resuscitative metric in hemorrhagic shock.Trauma Surg Acute Care Open. 2024 Aug 24;9(1):e001480. doi: 10.1136/tsaco-2024-001480. eCollection 2024. Trauma Surg Acute Care Open. 2024. PMID: 39296600 Free PMC article.
-
Pulse oximetry: fundamentals and technology update.Med Devices (Auckl). 2014 Jul 8;7:231-9. doi: 10.2147/MDER.S47319. eCollection 2014. Med Devices (Auckl). 2014. PMID: 25031547 Free PMC article. Review.
-
A multicentre prospective observational study comparing arterial blood gas values to those obtained by pulse oximeters used in adult patients attending Australian and New Zealand hospitals.BMC Pulm Med. 2020 Jan 9;20(1):7. doi: 10.1186/s12890-019-1007-3. BMC Pulm Med. 2020. PMID: 31918697 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources