Abstract
Background: In 2002, the National Institute of Standards and Technology (NIST) established a reference method for serum potassium based on inductively coupled plasma-mass spectrometry (ICP-MS). The aim of this study was to develop an inexpensive and improved candidate reference method for accurate and precise determination of serum potassium.
Methods: Serum samples were diluted with 1% HNO3 supplemented with (59)Cobalt as isotope internal standard, and potassium was measured by ICP-MS. The new method was evaluated with NIST standard reference materials (SRMs), according to the Clinical and Laboratory Standard Institute's evaluation protocols.
Results: At 4.300 and 4.678 mmol/l levels, the present method demonstrated analytical imprecision of 0.09% and 0.14%, and recoveries of 99.67% to 99.88%, respectively. The bias between the target values of SRMs were -0.02% to +0.28%, respectively. This method was linear between 0.0000 and 6.87 mmol/l (R(2)=1.000). The method had an uncertainty (U95%) of 0.76%.
Conclusions: The proposed ICP-MS method to measure serum potassium is precise and accurate, with high sensitivity and specificity. It may be considered as a candidate reference method for the determination of serum potassium.
Copyright © 2012 Elsevier B.V. All rights reserved.
Similar articles
-
A candidate reference method for serum potassium measurement by inductively coupled plasma mass spectrometry.Clin Chem Lab Med. 2017 Aug 28;55(10):1517-1522. doi: 10.1515/cclm-2016-0803. Clin Chem Lab Med. 2017. PMID: 28361780
-
A candidate reference method for serum calcium measurement by inductively coupled plasma mass spectrometry.Clin Chim Acta. 2016 Oct 1;461:141-5. doi: 10.1016/j.cca.2016.07.004. Epub 2016 Jul 8. Clin Chim Acta. 2016. PMID: 27402592
-
Determination of serum calcium levels by 42Ca isotope dilution inductively coupled plasma mass spectrometry.Clin Chem Lab Med. 2017 Nov 27;56(1):51-58. doi: 10.1515/cclm-2017-0175. Clin Chem Lab Med. 2017. PMID: 28704178
-
Isotope ratio determination in boron analysis.Biol Trace Elem Res. 1998 Winter;66(1-3):39-53. doi: 10.1007/BF02783124. Biol Trace Elem Res. 1998. PMID: 10050906 Review.
-
A review of the capabilities of ICP-MS for trace element analysis in body fluids and tissues.J Trace Elem Electrolytes Health Dis. 1993 Sep;7(3):131-9. J Trace Elem Electrolytes Health Dis. 1993. PMID: 8155984 Review.
Cited by
-
An improved reference method for serum cations measurement by ion chromatography.J Clin Lab Anal. 2018 Jul;32(6):e22429. doi: 10.1002/jcla.22429. Epub 2018 Mar 30. J Clin Lab Anal. 2018. PMID: 29603394 Free PMC article.
-
Stable sensing platform for diagnosing electrolyte disturbance using laser-induced breakdown spectroscopy.Biomed Opt Express. 2022 Dec 1;13(12):6778-6790. doi: 10.1364/BOE.477565. eCollection 2022 Dec 1. Biomed Opt Express. 2022. PMID: 36589579 Free PMC article.
-