Transillumination imaging performance: spatial resolution simulation studies
- PMID: 2308546
- DOI: 10.1118/1.596526
Transillumination imaging performance: spatial resolution simulation studies
Abstract
Monte Carlo calculations were used to simulate the propagation of visible and near-infrared light through homogenous tissue in order to quantitate the potential spatial resolution performance for transillumination imaging. Specifically, the relative effectiveness of coaxial collimation and time of flight (TOF) detection for improving spatial resolution was investigated. The results demonstrate that significant improvements in spatial resolution can be achieved through these techniques, with TOF methods offering superior performance for a given level of detected signal intensity.
Similar articles
-
Transillumination imaging performance: a time-of-flight imaging system.Med Phys. 1990 May-Jun;17(3):351-6. doi: 10.1118/1.596514. Med Phys. 1990. PMID: 2385191
-
Spatial resolution in fast time-resolved transillumination imaging: an indeterministic Monte Carlo approach.Phys Med Biol. 2002 Aug 21;47(16):2935-45. doi: 10.1088/0031-9155/47/16/309. Phys Med Biol. 2002. PMID: 12222857
-
Monte Carlo modelling of light propagation in breast tissue.Phys Med Biol. 1991 May;36(5):591-602. doi: 10.1088/0031-9155/36/5/003. Phys Med Biol. 1991. PMID: 2068225
-
Transillumination imaging for blood oxygen saturation estimation of skin lesions.IEEE Trans Biomed Eng. 2012 Sep;59(9):2660-7. doi: 10.1109/TBME.2012.2209647. Epub 2012 Jul 20. IEEE Trans Biomed Eng. 2012. PMID: 22835531
-
Near-Infrared Transillumination for Macroscopic Functional Imaging of Animal Bodies.Biology (Basel). 2023 Oct 24;12(11):1362. doi: 10.3390/biology12111362. Biology (Basel). 2023. PMID: 37997961 Free PMC article. Review.
Cited by
-
Laser tomography of heterogeneous scattering media using spatial and temporal resolution.Med Biol Eng Comput. 1993 Mar;31(2):157-64. doi: 10.1007/BF02446674. Med Biol Eng Comput. 1993. PMID: 8331996
-
Simulation of time-resolved breast transillumination.Med Biol Eng Comput. 1993 Mar;31(2):165-70. doi: 10.1007/BF02446675. Med Biol Eng Comput. 1993. PMID: 8331997
-
Transcranial optical path length in infants by near-infrared phase-shift spectroscopy.J Clin Monit. 1995 Mar;11(2):109-17. doi: 10.1007/BF01617732. J Clin Monit. 1995. PMID: 7760083
-
Diagnosis of defects in oxidative muscle metabolism by non-invasive tissue oximetry.Mol Cell Biochem. 1997 Sep;174(1-2):7-10. Mol Cell Biochem. 1997. PMID: 9309658
MeSH terms
LinkOut - more resources
Full Text Sources