Comparative Study
. 2013 Apr;23(4):1015-26.
doi: 10.1007/s00330-012-2686-9. Epub 2012 Oct 20.

Differentiation of benign from malignant solid breast masses: comparison of two-dimensional and three-dimensional shear-wave elastography

Affiliations
Comparative Study

Differentiation of benign from malignant solid breast masses: comparison of two-dimensional and three-dimensional shear-wave elastography

Su Hyun Lee et al. Eur Radiol. 2013 Apr.

Abstract

Objectives: To prospectively compare the diagnostic performances of two-dimensional (2D) and three-dimensional (3D) shear-wave elastography (SWE) for differentiating benign from malignant breast masses.

Methods: B-mode ultrasound and SWE were performed for 134 consecutive women with 144 breast masses before biopsy. Quantitative elasticity values (maximum and mean elasticity in the stiffest portion of mass, Emax and Emean; lesion-to-fat elasticity ratio, Erat) were measured with both 2D and 3D SWE. The area under the receiver operating characteristic curve (AUC), sensitivity and specificity of B-mode, 2D, 3D SWE and combined data of B-mode and SWE were compared.

Results: Sixty-seven of the 144 breast masses (47 %) were malignant. Overall, higher elasticity values of 3D SWE than 2D SWE were noted for both benign and malignant masses. The AUC for 2D and 3D SWE were not significantly different: Emean, 0.938 vs 0.928; Emax, 0.939 vs 0.930; Erat, 0.907 vs 0.871. Either 2D or 3D SWE significantly improved the specificity of B-mode ultrasound from 29.9 % (23 of 77) up to 71.4 % (55 of 77) and 63.6 % (49 of 77) without a significant change in sensitivity.

Conclusion: Two-dimensional and 3D SWE performed equally in distinguishing benign from malignant masses and both techniques improved the specificity of B-mode ultrasound.

PubMed Disclaimer

Similar articles

Cited by

References

    1. N Engl J Med. 2012 Mar 8;366(10):956-7 - PubMed
    1. Breast Cancer Res Treat. 2011 Aug;129(1):89-97 - PubMed
    1. Radiology. 2011 Apr;259(1):59-64 - PubMed
    1. Radiol Clin North Am. 2004 Sep;42(5):845-51, vi - PubMed
    1. Radiology. 2012 Jun;263(3):673-7 - PubMed

Publication types

LinkOut - more resources