External quality assurance performance of clinical research laboratories in sub-saharan Africa
- PMID: 23086773
- DOI: 10.1309/AJCP8PCM4JVLEEQR
External quality assurance performance of clinical research laboratories in sub-saharan Africa
Abstract
Patient Safety Monitoring in International Laboratories (JHU-SMILE) is a resource at Johns Hopkins University that supports and monitors laboratories in National Institutes of Health-funded international clinical trials. To determine the impact of the JHU-SMILE quality assurance scheme in sub-Saharan African laboratories, we reviewed 40 to 60 months of external quality assurance (EQA) results of the College of American Pathologists (CAP) in these laboratories. We reviewed the performance of 8 analytes: albumin, alanine aminotransferase, creatinine, sodium, WBC, hemoglobin, hematocrit, and the human immunodeficiency virus antibody rapid test. Over the 40- to 60-month observation period, the sub-Saharan laboratories had a 1.63% failure rate, which was 40% lower than the 2011 CAP-wide rate of 2.8%. Seventy-six percent of the observed EQA failures occurred in 4 of the 21 laboratories. These results demonstrate that a system of remote monitoring, feedback, and audits can support quality in low-resource settings, even in places without strong regulatory support for laboratory quality.
Comment in
-
Laboratory quality control for all. Can this be achieved?Am J Clin Pathol. 2012 Nov;138(5):639-41. doi: 10.1309/AJCP77JZIJAGUIDR. Am J Clin Pathol. 2012. PMID: 23086763 No abstract available.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Miscellaneous