. 2013 Jun;118(4):540-54.
doi: 10.1007/s11547-012-0892-8. Epub 2012 Oct 22.

Optimisation of radiological protocols for chest imaging using computed radiography and flat-panel X-ray detectors

Affiliations

Optimisation of radiological protocols for chest imaging using computed radiography and flat-panel X-ray detectors

G Compagnone et al. Radiol Med. 2013 Jun.

Abstract

Purpose: Digital radiography technology has replaced conventional screen-film systems in many hospitals. Despite the different characteristics of new detector materials, frequently, the same radiological protocols previously optimised for screen film are still used with digital equipment without any critical review. This study addressed optimisation of exposure settings for chest examinations with digital systems, considering both image quality and patient dose.

Materials and methods: Images acquired with direct digital radiography equipment and a computed radiography system were analysed with specially developed commercial software with a four-alternative forced-choice method: the most promising protocols were then scored by two senior radiologists.

Results: Digital technology offers a wide dynamic range and the ability to postprocess images, allowing use of lower tube potentials in chest examinations. The computed radiography system showed both better image quality and lower dose at lower energies (85 kVp and 95 kVp) than those currently used (125 kVp). Direct digital radiography equipment confirmed both its superior image quality and lower dose requirements compared with the storage phosphor plate system.

Conclusions: Generally, lowering tube potentials in chest examinations seems to allow better image quality/effective dose ratio when using digital equipment.

PubMed Disclaimer

References

    1. Radiat Prot Dosimetry. 2008;129(1-3):140-3 - PubMed
    1. Phys Med Biol. 2000 Sep;45(9):2653-69 - PubMed
    1. Radiology. 2000 Dec;217(3):707-12 - PubMed
    1. Med Phys. 2001 Nov;28(11):2328-35 - PubMed
    1. AJR Am J Roentgenol. 2006 Sep;187(3):630-7 - PubMed