Investigation of American Association of Physicists in Medicine Report 204 size-specific dose estimates for pediatric CT implementation
- PMID: 23093679
- DOI: 10.1148/radiol.12120131
Investigation of American Association of Physicists in Medicine Report 204 size-specific dose estimates for pediatric CT implementation
Abstract
Purpose: To compare five methodologies the American Association of Physicists in Medicine Report 204 used to calculate size-specific dose estimates (SSDEs) for pediatric computed tomography (CT).
Materials and methods: The institutional review board waived consent for this HIPAA-compliant retrospective study. The five SSDE methodologies were investigated for calculation variation based on volumetric CT dose index (CTDI), or CTDI(vol), of chest, abdominal, and pelvic CT. SSDE calculations were derived from a predominantly pediatric population of 186 patients retrospectively and consecutively analyzed from June through November 2011. Eighty (43%) of the 186 patients were female, and 106 (57%) were male. Mean patient age was 8.6 years ± 6.3 (standard deviation), the age range was 1 month to 28 years, and mean weight was 37.7 kg ± 33.1, with a range of 3.4-146.6 kg. SSDE conversion factors were derived from anteroposterior (AP) and lateral dimensions measured on the patient's CT radiograph. The measurements were either used independently, or as a summation, or to calculate the patient's effective diameter; additionally, SSDE was derived on the basis of the patient's age (International Commission on Radiation Units Report 74 data). SSDE conversion factors were applied to CTDI(vol) data that corrected for both 16- and 32-cm-diameter CTDI phantom measurements. SSDE data were summarized by using the patient's originally prescribed weight-based CT scanning protocols. Data were summarized by using descriptive statistics.
Results: SSDEs derived from individual measurements varied 2%-12%. The combination of measurements (sum or effective diameter) varied 0.9%-2%. The age approach varied by an average of 2% (in the younger population [0-13 years]), but up to 44%, with an average of 18% (in the older population [14-18 years]). No SSDE correction was required for patients of varying size who weighed 36 kg or less when CTDI(vol) was measured by using a 16-cm CTDI phantom or for patients weighing 100-140 kg when CTDI(vol) was measured by using a 32-cm phantom. CTDI(vol) measured by using a 32-cm phantom in patients weighing between 36 and 100 kg and patients weighing more than 140 kg differed from SSDE by an average of 35%. An average difference of 1% was found between male and female SSDE-corrected values when the two sexes were compared within the same CT weight scanning categories.
Conclusion: The combination of AP and lateral measurements should be used to determine SSDE correction factors when possible. For pediatric patients, CTDI(vol) calculated with a 32-cm phantom requires SSDE conversion to more accurately estimate patient dose; CTDI(vol) calculated with a 16-cm phantom for pediatric patients weighing 36 kg or less does not require SSDE conversion.
© RSNA, 2012.
Comment in
-
Size-specific dose estimation for CT: how should it be used and what does it mean?Radiology. 2012 Dec;265(3):666-8. doi: 10.1148/radiol.12121919. Radiology. 2012. PMID: 23175540 No abstract available.
Similar articles
-
Size-specific dose estimates for adult patients at CT of the torso.Radiology. 2012 Dec;265(3):841-7. doi: 10.1148/radiol.12112365. Epub 2012 Oct 22. Radiology. 2012. PMID: 23091173
-
Entrance skin dosimetry and size-specific dose estimate from pediatric chest CTA.J Cardiovasc Comput Tomogr. 2014 Mar-Apr;8(2):97-107. doi: 10.1016/j.jcct.2013.08.002. Epub 2013 Aug 23. J Cardiovasc Comput Tomogr. 2014. PMID: 24211194
-
Diagnostic reference ranges for pediatric abdominal CT.Radiology. 2013 Jul;268(1):208-18. doi: 10.1148/radiol.13120730. Epub 2013 Mar 19. Radiology. 2013. PMID: 23513245
-
Estimated pediatric radiation dose during CT.Pediatr Radiol. 2011 Sep;41 Suppl 2:472-82. doi: 10.1007/s00247-011-2179-z. Epub 2011 Aug 17. Pediatr Radiol. 2011. PMID: 21847725 Review.
-
Dose indices: everybody wants a number.Pediatr Radiol. 2014 Oct;44 Suppl 3:450-9. doi: 10.1007/s00247-014-3104-z. Epub 2014 Oct 11. Pediatr Radiol. 2014. PMID: 25304704 Review.
Cited by
-
Development and validation of an open source Monte Carlo dosimetry model for wide-beam CT scanners using Fluka.J Appl Clin Med Phys. 2019 Apr;20(4):132-147. doi: 10.1002/acm2.12559. Epub 2019 Mar 9. J Appl Clin Med Phys. 2019. PMID: 30851155 Free PMC article.
-
Establishment of local diagnostic reference levels for computed tomography with cloud-based automated dose-tracking software in Türkiye.Diagn Interv Radiol. 2024 May 13;30(3):205-211. doi: 10.4274/dir.2023.232265. Epub 2023 Aug 31. Diagn Interv Radiol. 2024. PMID: 37650522 Free PMC article.
-
Deep learning reconstruction improves image quality of abdominal ultra-high-resolution CT.Eur Radiol. 2019 Nov;29(11):6163-6171. doi: 10.1007/s00330-019-06170-3. Epub 2019 Apr 11. Eur Radiol. 2019. PMID: 30976831
-
Asian consortium on radiation dose of pediatric cardiac CT (ASCI-REDCARD).Pediatr Radiol. 2017 Jul;47(8):899-910. doi: 10.1007/s00247-017-3847-4. Epub 2017 Apr 24. Pediatr Radiol. 2017. PMID: 28435986
-
Applying the AAPM 293 report to estimate the absorbed dose during head computed tomography: using head circumference for rapid dose estimation.Quant Imaging Med Surg. 2023 May 1;13(5):3140-3149. doi: 10.21037/qims-22-983. Epub 2023 Mar 10. Quant Imaging Med Surg. 2023. PMID: 37179955 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources