Physiologic trend detection and artifact rejection: a parallel implementation of a multi-state Kalman filtering algorithm
- PMID: 2311364
- DOI: 10.1016/0169-2607(90)90026-6
Physiologic trend detection and artifact rejection: a parallel implementation of a multi-state Kalman filtering algorithm
Abstract
Using a parallel implementation of the multi-state Kalman filtering algorithm, we have developed an accurate method of reliably detecting and identifying trends, abrupt changes, and artifacts from multiple physiologic data streams in real-time. The Kalman filter algorithm was implemented within an innovative software architecture for parallel computation: a parallel process trellis. Examples, processed in real-time, of both simulated and actual data serve to illustrate the potential value of the Kalman filter as a tool in physiologic monitoring.