Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2013 Feb;16(2):129-33.
doi: 10.1093/icvts/ivs457. Epub 2012 Nov 1.

Comparison of radial deformability of stent posts of different aortic bioprostheses

Affiliations
Comparative Study

Comparison of radial deformability of stent posts of different aortic bioprostheses

Martins Kalejs et al. Interact Cardiovasc Thorac Surg. 2013 Feb.

Abstract

Objectives: Little is known about the stent deformability required for optimal stented heart valve bioprosthesis design. Therefore, two bioprosthetic valves with known good long-term clinical results were tested. The strain in the radial direction of the stent posts of these valves was compared with contemporary bioprosthetic valves and a native porcine aortic root.

Methods: Medtronic Intact and Carpentier-Edwards Standard (CES), and four contemporary bioprostheses, including one self-expanding prosthesis, were tested with three sonomicrometry probes per valve fixed at commissure attachment points. The mean values from 2400 data points from three measurements of the interprobe distances were used to calculate the radius of the circle circumscribed around the three probes. Changes in the radius of the aortic root at pressures 70-90 and 120-140 mmHg (pressure during diastole and systole) and that of the stent posts at 70-90 and 0-10 mmHg (transvalvular pressure gradient during diastole and systole) were compared.

Results: An increase in radius by 7.3 ± 2.6, 8.7 ± 0.0 and 3.9 ± 0.0% for the porcine aortic root, CES and Intact valves, respectively, was observed during transition from diastolic to systolic pressure and less for contemporary bioprostheses-mean 2.5 ± 0.9%, lowest 1.2 ± 0.0.

Conclusions: The results indicate that the radial deformability of bioprosthetic valve stent posts can be as low as 1.2% for xenoaortic and 3.0% for xenopericardial prostheses with no compromise of valve durability. Although these results suggest that valve stent post-deformability might not be of critical importance, a concrete answer to the question of the significance of stent deformability for valve durability can be obtained only by acquiring long-term follow-up results for valve prostheses with rigid stents.

PubMed Disclaimer

Figures

Figure 1:
Figure 1:
Photographs of bioprostheses tested: (a) St. Jude Epic, (b) Medtronic Hancock II, (c) Sorin Soprano, (d) ATS 3F Enable, (e) Medtronic Intact and (f) Carpentier-Edwards Standard.
Figure 2:
Figure 2:
A schematic representation of the test set-up used in this experiment. (1) Valve-mounting chamber; (2) chamber with two ports; (3) water bath with 0.9% NaCl solution at room temperature; (4) roller pump; (5) system for real-time pressure recording; (6) endoscope; (7) endoscope ‘tower’; (8) three Sonometrics probes attached at the inner sides of stent posts.

References

    1. Vesely I. The evolution of bioprosthetic heart valve design and its impact on durability. Cardiovasc Pathol. 2003;12:277–86. - PubMed
    1. Christie GW. Computer modeling of bioprosthetic heart valves. Eur J Cardiothorac Surg. 2003;6:S95–S101. - PubMed
    1. Jamieson WR, Miyagishima RT, Burr LH, Lichtenstein SV, Fradet GJ, Janusz MT. Carpentier-Edwards porcine bioprostheses: clinical performance assessed by actual analysis. J Heart Valve Dis. 2000;9:530–5. - PubMed
    1. de la Fuente A, Sánchez R, Imizcoz A, Fernández JL, Olaz F, Moriones I. Intact Medtronic and Carpentier Edwards S.A.V. clinical and hemodynamic outcomes over 13 years. Cardiovasc Surg. 2003;11:139–44. - PubMed
    1. de la Fuente A, Sánchez JR, Fernández JL, Romero J, Berjon J, Moriones I. The Medtronic Intact bioprosthesis: clinical and hemodynamic performance over 13 years. Artif Organs. 2002;26:851–5. - PubMed

Publication types