Mid-term results of small-sized St. Jude Medical Regent prosthetic valves (21 mm or less) for small aortic annulus
- PMID: 23160858
- DOI: 10.1007/s00380-012-0306-x
Mid-term results of small-sized St. Jude Medical Regent prosthetic valves (21 mm or less) for small aortic annulus
Abstract
Prosthesis-patient mismatch (PPM) is always of concern when performing aortic valve replacement (AVR) in patients with a small aortic annulus. Although bioprosthetic AVR is preferred in patients older than 65 years, we have experienced cases in elderly patients with a small aortic annulus whereby we could not implant small-sized bioprosthetic valves. We have implanted St. Jude Medical Regent (SJMR) mechanical valves (St. Jude Medical, St. Paul, MN, USA) as necessary, even in elderly patients with no aortic annulus enlargement. We investigated our experiences of AVR with SJMR mechanical valves of 21 mm or less in size. Between January 2006 and December 2009, 40 patients underwent AVR with SJMR mechanical valves ≤21 mm in size: 9 patients received 21-mm valves, 19 received 19-mm valves, and 12 received 17-mm valves. The mean age was 65.9 ± 9.5 years, and 25 patients (62.5 %) were 65 years or older. We evaluated the clinical outcome and the echocardiographic data for each valve size. There was no operative or hospital mortality. The mean duration of clinical follow-up was 31.2 ± 17.6 months. During follow-up, there were no hospitalizations due to heart failure. The cumulative valve-related event-free survival was 93 % at 33 months, and the cumulative hemorrhagic event-free survival was 93 % at 33 months and 84 % at 43 months, using the Kaplan-Meier method. At follow-up, the mean values of the measured effective orifice area (EOA) for the 21-, 19-, and 17-mm prostheses were 2.00 ± 0.22, 1.74 ± 0.37, and 1.25 ± 0.26 cm(2), and the mean measured EOA index (EOAI) were 1.17 ± 0.12, 1.11 ± 0.21 and 0.90 ± 0.22 cm(2)/m(2), respectively. A PPM (EOAI ≤0.85) was documented in 5 patients, all of whom had received a 17-mm SJMR valve. AVR with SJMR valves of 21 mm or less in size appears to show satisfactory clinical and hemodynamic results.
Similar articles
-
Early and Mid-term Outcome of the St. Jude Medical Regent 19-mm Aortic Valve Mechanical Prosthesis. Functional and Haemodynamic Evaluation.Heart Lung Circ. 2018 Feb;27(2):235-247. doi: 10.1016/j.hlc.2017.02.017. Epub 2017 Mar 28. Heart Lung Circ. 2018. PMID: 28400190
-
Aortic valve replacement with 17-mm St. Jude Medical Regent prosthetic valves for a small calcified aortic annulus in elderly patients.Gen Thorac Cardiovasc Surg. 2010 Oct;58(10):506-10. doi: 10.1007/s11748-010-0629-4. Epub 2010 Oct 13. Gen Thorac Cardiovasc Surg. 2010. PMID: 20941563
-
Comprehensive hemodynamic performance and frequency of patient-prosthesis mismatch of the St. Jude Medical Trifecta bioprosthetic aortic valve.J Heart Valve Dis. 2014 Jul;23(4):516-23. J Heart Valve Dis. 2014. PMID: 25803979
-
Assessing options for the small aortic root.J Heart Valve Dis. 2002 Jan;11 Suppl 1:S50-5. J Heart Valve Dis. 2002. PMID: 11843521 Review.
-
Stentless aortic valve replacement: an update.Vasc Health Risk Manag. 2011;7:345-51. doi: 10.2147/VHRM.S11253. Epub 2011 Jun 2. Vasc Health Risk Manag. 2011. PMID: 21731886 Free PMC article. Review.
Cited by
-
Prosthesis-patient mismatch - what cardiac anesthesiologists need to know?Ann Card Anaesth. 2017 Apr-Jun;20(2):234-242. doi: 10.4103/aca.ACA_9_17. Ann Card Anaesth. 2017. PMID: 28393786 Free PMC article. Review.
-
Late clinical outcomes after mechanical aortic valve replacement for aortic stenosis: old versus new prostheses.J Thorac Dis. 2018 Jun;10(6):3361-3371. doi: 10.21037/jtd.2018.06.03. J Thorac Dis. 2018. PMID: 30069331 Free PMC article.
-
Early outcomes of the "Chimney" commando procedure in the small aortic and mitral annuli.Front Cardiovasc Med. 2023 Jul 24;10:1139771. doi: 10.3389/fcvm.2023.1139771. eCollection 2023. Front Cardiovasc Med. 2023. PMID: 37554367 Free PMC article.
-
Seventeen-millimeter St. Jude Medical Regent valve in patients with small aortic annulus: dose moderate prosthesis-patient mismatch matter?J Cardiothorac Surg. 2014 Jan 17;9:17. doi: 10.1186/1749-8090-9-17. J Cardiothorac Surg. 2014. PMID: 24438101 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical