Abstract
Pulse oximetry is a standard noninvasive procedure for monitoring arterial oxygen saturation. Finger injuries related to pulse oximeter use have been reported as chemical or thermal burns, sun-tanning, pressure erosion, sensory loss, and gangrene. The mechanisms of these complications have not been definitively explained; but pressure ischemia, prolonged use, overheating of the probe, inappropriate use of the probe, and short circuiting are considered to be the main factors. We describe 2 cases of pulse oximeter probe-induced finger injuries, propose the possible mechanisms and factors related to the injury, and discuss the management.
Copyright © 2012 Elsevier Inc. All rights reserved.
Similar articles
-
[A case report of a burn produced by the probe of a pulse oximeter].Masui. 1992 Dec;41(12):1991-3. Masui. 1992. PMID: 1479670 Japanese.
-
Pulse oximeter-associated toe injuries in a premature neonate: a case report.Zhonghua Yi Xue Za Zhi (Taipei). 1999 Dec;62(12):914-6. Zhonghua Yi Xue Za Zhi (Taipei). 1999. PMID: 10634008
-
Finger injury resulting from pulse oximeter probe during orthognathic surgery.Oral Surg Oral Med Oral Pathol. 1990 Feb;69(2):161-3. doi: 10.1016/0030-4220(90)90318-m. Oral Surg Oral Med Oral Pathol. 1990. PMID: 2304740
-
Pulse oximetry in the oesophagus.Physiol Meas. 2006 Jan;27(1):R1-35. doi: 10.1088/0967-3334/27/1/R01. Epub 2005 Nov 29. Physiol Meas. 2006. PMID: 16365505 Review.
-
Pulse oximetry: a review of the theory, accuracy, and clinical applications.Obstet Gynecol. 1989 Sep;74(3 Pt 2):541-6. Obstet Gynecol. 1989. PMID: 2668828 Review.
Cited by
-
Second-Degree Burns in Neonates: A Rare Case Report of Saturation Probe Injury in Neonates.Cureus. 2023 Oct 26;15(10):e47761. doi: 10.7759/cureus.47761. eCollection 2023 Oct. Cureus. 2023. PMID: 38022296 Free PMC article.
-
A case of a hand burns associated with a "wrap-around" pulse oximeter in an infant.Int Wound J. 2020 Aug;17(4):1097-1098. doi: 10.1111/iwj.13375. Epub 2020 Apr 25. Int Wound J. 2020. PMID: 32333718 Free PMC article. No abstract available.
-
Reflective oxygen saturation monitoring at hypothenar and its validation by human hypoxia experiment.Biomed Eng Online. 2015 Aug 5;14:76. doi: 10.1186/s12938-015-0071-z. Biomed Eng Online. 2015. PMID: 26242309 Free PMC article.
-
Skin-interfacing wearable biosensors for smart health monitoring of infants and neonates.Commun Mater. 2024;5(1):72. doi: 10.1038/s43246-024-00511-6. Epub 2024 May 9. Commun Mater. 2024. PMID: 38737724 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical