Measurement accuracy of heart rate and respiratory rate during graded exercise and sustained exercise in the heat using the Zephyr BioHarness
- PMID: 23175181
- PMCID: PMC4620538
- DOI: 10.1055/s-0032-1327661
Measurement accuracy of heart rate and respiratory rate during graded exercise and sustained exercise in the heat using the Zephyr BioHarness
Abstract
The Zephyr BioHarness was tested to determine the accuracy of heart rate (HR) and respiratory rate (RR) measurements during 2 exercise protocols in conjunction with either a laboratory metabolic cart (Vmax) or a previously validated portable metabolic system (K4b2). In one protocol, HR and RR were measured using the BioHarness and Vmax during a graded exercise up to V˙O2max (n=12). In another protocol, HR and RR were measured using the BH and K4b2 during sustained exercise (30% and 50% V˙O2max for 20 min each) in a hot environment (30 °C, 50% relative humidity) (n=6). During the graded exercise, HR but not RR, obtained from the BioHarness was higher compared to the Vmax at baseline and 30% V˙O2max (p<0.05), but showed no significant difference at other stages with high correlation coefficients for both HR (r=0.87-0.96) and RR (r=0.90-0.99 above 30% V˙O2max). During the exercise in the heat, there were no significant differences between the BioHarness and K4b2 system. Correlation coefficients between the methods were low for HR but moderately to highly correlated (0.49-0.99) for RR. In conclusion, the BioHarness is comparable to Vmax and K4b2 over a wide range of V˙O2 during graded exercise and sustained exercise in the heat.
© Georg Thieme Verlag KG Stuttgart · New York.
Conflict of interest statement
Figures
Similar articles
-
Defining the number of bouts and oxygen uptake during the "Tabata protocol" performed at different intensities.Physiol Behav. 2018 May 15;189:10-15. doi: 10.1016/j.physbeh.2018.02.045. Epub 2018 Feb 24. Physiol Behav. 2018. PMID: 29486169
-
Reliability of Zephyr Bioharness and Fitbit Charge Measures of Heart Rate and Activity at Rest, During the Modified Canadian Aerobic Fitness Test, and Recovery.J Strength Cond Res. 2019 Feb;33(2):559-571. doi: 10.1519/JSC.0000000000001842. J Strength Cond Res. 2019. PMID: 30689619
-
Heat Acclimation with Controlled Heart Rate: Influence of Hydration Status.Med Sci Sports Exerc. 2020 Aug;52(8):1815-1824. doi: 10.1249/MSS.0000000000002320. Med Sci Sports Exerc. 2020. PMID: 32102056
-
Reliability of Zephyr BioHarness Respiratory Rate at Rest, During the Modified Canadian Aerobic Fitness Test and Recovery.J Strength Cond Res. 2020 Jan;34(1):264-269. doi: 10.1519/JSC.0000000000003046. J Strength Cond Res. 2020. PMID: 30694966
-
A new objective method for determining exercise gas exchange thresholds by respiratory frequency in middle-aged men.Eur J Appl Physiol. 2024 Nov;124(11):3227-3240. doi: 10.1007/s00421-024-05520-4. Epub 2024 Jun 7. Eur J Appl Physiol. 2024. PMID: 38849689 Free PMC article.
Cited by
-
Remote Monitoring of Chronic Critically Ill Patients after Hospital Discharge: A Systematic Review.J Clin Med. 2022 Feb 15;11(4):1010. doi: 10.3390/jcm11041010. J Clin Med. 2022. PMID: 35207287 Free PMC article. Review.
-
Estimating energy expenditure of head-hauling water and grain grinding from heart rate monitor measurements in northern Mozambique.Public Health Nutr. 2020 Nov;23(16):2886-2897. doi: 10.1017/S1368980020001019. Epub 2020 Jul 2. Public Health Nutr. 2020. PMID: 32611478 Free PMC article.
-
Inter-Instrument Reliability and Agreement of Fitbit Charge Measurements of Heart Rate and Activity at Rest, during the Modified Canadian Aerobic Fitness Test, and in Recovery.Physiother Can. 2019 Summer;71(3):197-206. doi: 10.3138/ptc.2018-25. Physiother Can. 2019. PMID: 31719715 Free PMC article.
-
Wearable sensors for monitoring the internal and external workload of the athlete.NPJ Digit Med. 2019 Jul 29;2:71. doi: 10.1038/s41746-019-0149-2. eCollection 2019. NPJ Digit Med. 2019. PMID: 31372506 Free PMC article. Review.
-
Remote Photoplethysmography Is an Accurate Method to Remotely Measure Respiratory Rate: A Hospital-Based Trial.J Clin Med. 2022 Jun 24;11(13):3647. doi: 10.3390/jcm11133647. J Clin Med. 2022. PMID: 35806932 Free PMC article.
References
-
- Adams W. Measurement of breathing rate and volume in routinely performed daily activities. [Accessed on November 14, 2011];California Environmental protection Agency Air Resource Board Research Division Contract No, A033-205. 1993 Available from: http://www,arb,ca,gov/research/apr/past/a033-205.pdf.
-
- Al-Khalidi F, Saatchi R, Burke D, Elphick H, Tan S. Respiration rate monitoring methods: A review. Ped Pulmonol. 2011;46:523–529. - PubMed
-
- Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999;8:135–160. - PubMed
-
- Caruana-Montaldo B, Gleeson K, Zwillich CW. The control of breathing in clinical practice. Chest. 2000;117:205–225. - PubMed
-
- Coca A, Roberge RJ, Williams WJ, Landsittel DP, Powell JB, Palmiero A. Physiological monitoring in firefighter ensembles: wearable plethysmographic sensor vest versus standard equipment. J Occup Environ Hyg. 2010;7:109–114. - PubMed
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical