Abstract
A rapid, enzymatic assay for serum or plasma paracetamol has been developed with the potential for adaptation to a wide range of clinical analysers. The method involves the action of an amidase enzyme to produce 4-aminophenol from paracetamol, which in turn reacts with 8-hydroxyquinoline in the presence of manganese ions to form a blue dye. Two stable reagents are used and excellent precision is achieved over the drug concentration range 0-2.5 mmol/l. The method, which is complete within 6 min, has been validated using a Monarch centrifugal analyser and shows no significant interference from endogenous serum compounds, drugs or paracetamol metabolites.