. 2013 Apr;36(3):307-14.
doi: 10.1007/s10840-012-9750-x. Epub 2012 Nov 21.

Clinical evaluation of a new technique to monitor return electrode skin temperature during radiofrequency ablation

Affiliations

Clinical evaluation of a new technique to monitor return electrode skin temperature during radiofrequency ablation

Siddharth J Trivedi et al. J Interv Card Electrophysiol. 2013 Apr.

Abstract

Purpose: Return electrode burns occur occasionally in cardiac radiofrequency ablation and more frequently in tumor radiofrequency ablation. A return electrode incorporating a thermochromic liquid crystal (TLC) layer, which changes color with temperature, has been shown in sheep studies to accurately indicate underlying skin temperature. We aimed to validate the accuracy of TLC-coated return electrodes in indicating skin temperature in the clinical setting of cardiac radiofrequency ablation.

Methods and results: The top layer of a standard return electrode was replaced with TLC. Fluoro-optic thermometer (FOT) probes were laid on the skin side of the return electrode, which was then placed on the left lateral mid-thigh of 18 patients (mean age = 61 ± 12 years, 12 men) undergoing cardiac radiofrequency ablation. Return electrode photographs were taken when FOT temperature exceeded 35 °C. TLC color changes, observed in 11 patients, were converted to temperature and compared with FOT temperature. TLC temperature correlated well with FOT temperature (Pearson's coefficient = 0.97 ± 0.03). Bland-Altman analysis showed good agreement (mean temperature difference = -0.04 ± 0.08 °C, upper limit of agreement = 0.11 ± 0.005 °C, lower limit of agreement = -0.19 ± 0.005 °C). The maximum FOT temperature recorded was 39.6 °C. There was no thermal injury at the return electrode site on any patients, when assessed immediately after and the day following the procedure.

Conclusion: TLC-coated return electrodes accurately indicate underlying skin temperature in cardiac radiofrequency ablation and may help prevent burns. This technology might be essential in high energy radiofrequency ablation.

PubMed Disclaimer

Similar articles

Cited by

References

    1. J Gastrointest Surg. 2001 Sep-Oct;5(5):477-89 - PubMed
    1. J Cardiovasc Electrophysiol. 2005 Jul;16(7):781-8 - PubMed
    1. Int J Hyperthermia. 2007 Nov;23(7):555-66 - PubMed
    1. AJR Am J Roentgenol. 2002 Apr;178(4):905-7 - PubMed
    1. Int J Hyperthermia. 2007 Jun;23(4):387-94 - PubMed

Publication types