Abstract
Purpose: Return electrode burns occur occasionally in cardiac radiofrequency ablation and more frequently in tumor radiofrequency ablation. A return electrode incorporating a thermochromic liquid crystal (TLC) layer, which changes color with temperature, has been shown in sheep studies to accurately indicate underlying skin temperature. We aimed to validate the accuracy of TLC-coated return electrodes in indicating skin temperature in the clinical setting of cardiac radiofrequency ablation.
Methods and results: The top layer of a standard return electrode was replaced with TLC. Fluoro-optic thermometer (FOT) probes were laid on the skin side of the return electrode, which was then placed on the left lateral mid-thigh of 18 patients (mean age = 61 ± 12 years, 12 men) undergoing cardiac radiofrequency ablation. Return electrode photographs were taken when FOT temperature exceeded 35 °C. TLC color changes, observed in 11 patients, were converted to temperature and compared with FOT temperature. TLC temperature correlated well with FOT temperature (Pearson's coefficient = 0.97 ± 0.03). Bland-Altman analysis showed good agreement (mean temperature difference = -0.04 ± 0.08 °C, upper limit of agreement = 0.11 ± 0.005 °C, lower limit of agreement = -0.19 ± 0.005 °C). The maximum FOT temperature recorded was 39.6 °C. There was no thermal injury at the return electrode site on any patients, when assessed immediately after and the day following the procedure.
Conclusion: TLC-coated return electrodes accurately indicate underlying skin temperature in cardiac radiofrequency ablation and may help prevent burns. This technology might be essential in high energy radiofrequency ablation.
Similar articles
-
A thermochromic dispersive electrode can measure the underlying skin temperature and prevent burns during radiofrequency ablation.J Cardiovasc Electrophysiol. 2005 Jul;16(7):781-8. doi: 10.1111/j.1540-8167.2005.40747.x. J Cardiovasc Electrophysiol. 2005. PMID: 16050838
-
Comparison of in vivo tissue temperature profile and lesion geometry for radiofrequency ablation with a saline-irrigated electrode versus temperature control in a canine thigh muscle preparation.Circulation. 1995 Apr 15;91(8):2264-73. doi: 10.1161/01.cir.91.8.2264. Circulation. 1995. PMID: 7697856
-
Skin burns associated with radiofrequency catheter ablation of cardiac arrhythmias.Pacing Clin Electrophysiol. 2013 Jun;36(6):764-7. doi: 10.1111/pace.12123. Epub 2013 Mar 19. Pacing Clin Electrophysiol. 2013. PMID: 23510131
-
Mechanisms of electrode induced injury. Part 3: practical concepts and avoidance.Am J Electroneurodiagnostic Technol. 2007 Dec;47(4):257-63. Am J Electroneurodiagnostic Technol. 2007. PMID: 18271314 Review.
-
Temperature monitoring during radiofrequency ablation.J Cardiovasc Electrophysiol. 1996 Feb;7(2):163-73. doi: 10.1111/j.1540-8167.1996.tb00511.x. J Cardiovasc Electrophysiol. 1996. PMID: 8853026 Review.
Cited by
-
Incidence and Factors Predicting Skin Burns at the Site of Indifferent Electrode during Radiofrequency Catheter Ablation of Cardiac Arrhythmias.Cardiol Res Pract. 2016;2016:5265682. doi: 10.1155/2016/5265682. Epub 2016 Apr 24. Cardiol Res Pract. 2016. PMID: 27213077 Free PMC article.