Comparative Study
. 2012 Nov-Dec;26(6):550-6.
doi: 10.1590/s1806-83242012000600011.

Bone density: comparative evaluation of Hounsfield units in multislice and cone-beam computed tomography

Affiliations
Free article
Comparative Study

Bone density: comparative evaluation of Hounsfield units in multislice and cone-beam computed tomography

Isabela Maria de Carvalho Crusoé Silva et al. Braz Oral Res. 2012 Nov-Dec.
Free article

Abstract

The aim of this study was to evaluate the validity of the bone density value of potential implant sites in HU obtained by a specific cone-beam computed tomography (CBCT) device. In this study, the HU values obtained using a MSCT scanner were used as the gold standard. Twenty mandibles (40 potential implant sites) were scanned using an MSCT scanner (Somatom Sensation 40) and a CBCT scanner (i-CAT). The MSCT images were evaluated using the Syngo CT Workplace software and the CBCT images, using the XoranCat software. The images were evaluated twice by three oral radiologists, at 60 day intervals. The trabecular bone density of the same area was evaluated on both images. Intraclass coefficients (ICC) were calculated to examine the agreement between the examiners and between the two periods of evaluation. The bone density and area of the ROI were compared by the Student t test and Bland-Altman analysis. ICCs were excellent. The mean HU value obtained using CBCT (418.06) was higher than that obtained using MSCT (313.13), with a statistically significant difference (p < 0.0001). In addition, Bland-Altman analysis showed that the HU measures were not equivalent. In conclusion, the bone density in HU with CBCT images obtained using the device studied proved unreliable, since it was higher than that obtained using MSCT.