Reliability of voxel gray values in cone beam computed tomography for preoperative implant planning assessment
- PMID: 23189294
Reliability of voxel gray values in cone beam computed tomography for preoperative implant planning assessment
Abstract
Purpose: To assess the reliability of cone beam computed tomography (CBCT) voxel gray value measurements using Hounsfield units (HU) derived from multislice computed tomography (MSCT) as a clinical reference (gold standard).
Materials and methods: Ten partially edentulous human mandibular cadavers were scanned by two types of computed tomography (CT) modalities: multislice CT and cone beam CT. On MSCT scans, eight regions of interest (ROI) designating the site for preoperative implant placement were selected in each mandible. The datasets from both CT systems were matched using a three-dimensional (3D) registration algorithm. The mean voxel gray values of the region around the implant sites were compared between MSCT and CBCT.
Results: Significant differences between the mean gray values obtained by CBCT and HU by MSCT were found. In all the selected ROIs, CBCT showed higher mean values than MSCT. A strong correlation (R=0.968) between mean voxel gray values of CBCT and mean HU of MSCT was determined.
Conclusions: Voxel gray values from CBCT deviate from actual HU units. However, a strong linear correlation exists, which may permit deriving actual HU units from CBCT using linear regression models.
Similar articles
-
Bone quality evaluation at dental implant site using multislice CT, micro-CT, and cone beam CT.Clin Oral Implants Res. 2015;26(1):e1-7. doi: 10.1111/clr.12315. Epub 2013 Dec 11. Clin Oral Implants Res. 2015. PMID: 24325572
-
Influence of cone beam CT scanning parameters on grey value measurements at an implant site.Dentomaxillofac Radiol. 2013;42(3):79884780. doi: 10.1259/dmfr/79884780. Epub 2012 Aug 29. Dentomaxillofac Radiol. 2013. PMID: 22933535 Free PMC article.
-
Comparison of gray values of cone-beam computed tomography with hounsfield units of multislice computed tomography: An in vitro study.Indian J Dent Res. 2017 Jan-Feb;28(1):66-70. doi: 10.4103/ijdr.IJDR_415_16. Indian J Dent Res. 2017. PMID: 28393820
-
Accuracy of linear measurements on CBCT images related to presurgical implant treatment planning: A systematic review.Clin Oral Implants Res. 2018 Oct;29 Suppl 16:393-415. doi: 10.1111/clr.13142. Clin Oral Implants Res. 2018. PMID: 30328204
-
CBCT-based bone quality assessment: are Hounsfield units applicable?Dentomaxillofac Radiol. 2015;44(1):20140238. doi: 10.1259/dmfr.20140238. Dentomaxillofac Radiol. 2015. PMID: 25315442 Free PMC article. Review.
Cited by
-
Comparison of Cone Beam Computed Tomography-Derived Alveolar Bone Density Between Subjects with and without Aggressive Periodontitis.J Clin Diagn Res. 2017 Jan;11(1):ZC118-ZC121. doi: 10.7860/JCDR/2017/22767.9305. Epub 2017 Jan 1. J Clin Diagn Res. 2017. PMID: 28274060 Free PMC article.
-
Feasibility of CBCT in Diagnosing Otosclerosis by Measuring CBCT Bone Density: A Preliminary Study.Indian J Otolaryngol Head Neck Surg. 2022 Sep;74(3):322-328. doi: 10.1007/s12070-021-02502-2. Epub 2021 Mar 26. Indian J Otolaryngol Head Neck Surg. 2022. PMID: 36213485 Free PMC article.
-
Associations among Primary Stability, Histomorphometric Findings, and Bone Density: A Prospective Randomized Study after Alveolar Ridge Preservation with a Collagen Cone.Dent J (Basel). 2020 Oct 2;8(4):112. doi: 10.3390/dj8040112. Dent J (Basel). 2020. PMID: 33023224 Free PMC article.
-
Comparison of the Hounsfield unit in CT scan with the gray level in cone-beam CT.J Dent Res Dent Clin Dent Prospects. 2019 Summer;13(3):177-182. doi: 10.15171/joddd.2019.028. Epub 2019 Oct 7. J Dent Res Dent Clin Dent Prospects. 2019. PMID: 31857863 Free PMC article.
-
Validation of bone mineral density measurement using quantitative CBCT image based on deep learning.Sci Rep. 2023 Jul 24;13(1):11921. doi: 10.1038/s41598-023-38943-8. Sci Rep. 2023. PMID: 37488135 Free PMC article.