. 1990 Jan;13(1):150-7.
doi: 10.1002/mrm.1910130114.

0.3-second FLASH MRI of the human heart

Affiliations

0.3-second FLASH MRI of the human heart

J Frahm et al. Magn Reson Med. 1990 Jan.

Abstract

Flow-suppressed FLASH MR images of the human heart have been recorded within a measuring time of 0.3 s using a 2.0-T whole-body research system (Siemens Magnetom) equipped with a conventional 10 mT m-1 gradient system. Subsecond imaging times have been achieved by reducing the repetition time to TR = 4.8 ms and by lowering the spatial resolution to 64 X 128 measured data points. The flip angle of the slice-selective radiofrequency (rf) pulses was adjusted to 10 degrees. Cardiac chambers, ventricular walls, and valves are well delineated in images from a single cardiac cycle using a field of 250 mm and a slice thickness of 8 mm. No motion artifacts were observed as a consequence of the short echo time of TE = 2.8 ms. Distinction between flowing blood and solid structures has been achieved by spatial presaturation of adjacent slices using two slice-selective 90 degrees rf pulses preceding the entire imaging sequence.