. 2013 Jan;34(1):115-20.
doi: 10.1097/MAO.0b013e318278522e.

Investigation of a novel completely-in-the-canal direct-drive hearing device: a temporal bone study

Affiliations

Investigation of a novel completely-in-the-canal direct-drive hearing device: a temporal bone study

Hossein Mahboubi et al. Otol Neurotol. 2013 Jan.

Abstract

Hypothesis: Whether a prototype direct-drive hearing device (DHD) is effective in driving the tympanic membrane (TM) in a temporal bone specimen to enable it to potentially treat moderate-to-severe hearing loss.

Background: Patient satisfaction with air conduction hearing aids has been low because of sound distortion, occlusion effect, and feedback issues. Implantable hearing aids provide a higher quality sound but require surgery for placement. The DHD was designed to combine the ability of driving the ossicular chain with placement in the external auditory canal.

Methods: DHD is a 3.5-mm wide device that could fit entirely into the bony ear canal and directly drive the TM rather than use a speaker. A cadaveric temporal bone was prepared. The device developed in our laboratory was coupled to the external surface of the TM and against the malleus. Frequency sweeps between 300 Hz to 12 kHz were performed in 2 different coupling methods at 104 and 120 dB, and the DHD was driven with various levels of current. Displacements of the posterior crus of the stapes were measured using a laser Doppler vibrometer.

Results: The DHD showed a linear frequency response from 300 Hz to 12 kHz. Placement against the malleus showed higher amplitudes and lower power requirements than when the device was placed on the TM.

Conclusion: DHD is a small completely-in-the-canal device that mechanically drives the TM. This novel device has a frequency output wider than most air conduction devices. Findings of the current study demonstrated that the DHD had the potential of being incorporated into a hearing aid in the future.

PubMed Disclaimer

Figures

Figure 1
Figure 1
The pioneer direct-drive hearing device that was tested on the cadaver temporal bone compared to a paper clip.
Figure 2
Figure 2
Schematic design of the direct-drive hearing device.
Figure 3
Figure 3
The temporal bone placed under direct illumination of the laser Doppler vibrometer. The reflective tape on the surface of the posterior crus is shown on the monitor.
Figure 4
Figure 4
Frequency response of the direct-drive hearing device at different voltage inputs when the device was coupled to the posterior-superior quadrant of the TM, compared to the background noise and the baseline measurements at 104 and 120 dB SPL.
Figure 5
Figure 5
Frequency response of the direct-drive hearing device at different voltage inputs when the device was coupled to the umbo, compared to the background noise and the baseline measurements at 104 and 120 dB SPL.

References

    1. Agrawal Y, Platz EA, Niparko JK. Risk factors for hearing loss in US adults: data from the National Health and Nutrition Examination Survey, 1999 to 2002. Otol Neurotol. 2009;30(2):139–145. - PubMed
    1. Ikeda N, Murray CJ, Salomon JA. Tracking population health based on self-reported impairments: Trends in the prevalence of hearing loss in US adults, 1976–2006. Am J Epidemiol. 2009;170(1):80–87. - PubMed
    1. Guidelines for the Audiological Management of Adult Hearing Impairment. Reston, VA: American Academy of Audiology; 2007. [Accessed January 20, 2012]. Available at: http://www.audiology.org/resources/documentlibrary/Documents/haguideline....
    1. Kochkin S. MarkeTrak VII: Hearing Loss Population Tops 31 Million People. The Hearing Review. 2005;12:16–29.
    1. Haynes DS, Young JA, Wanna GB, Glasscock ME., 3rd Middle ear implantable hearing devices: an overview. Trends Amplif. 2009;13(3):206–214. - PMC - PubMed

Publication types