Flanged-edge transverse gradient coil design for a hybrid LINAC-MRI system
- PMID: 23220182
- DOI: 10.1016/j.jmr.2012.11.017
Flanged-edge transverse gradient coil design for a hybrid LINAC-MRI system
Abstract
MRI can be combined with other systems, such as linear accelerators (LINACs) to provide image-guided therapy. However, in some configurations this requires splitting the MRI scanner to provide a central gap large enough to ensure dual access for the accelerator and the patient. This raises technical difficulties for maintaining a high gradient coil performance. In this research, a dedicated split transverse gradient coil was designed with a flange connected to the central coil end, which provided an additional surface for the current to flow. The coil was compared to existing designs, in terms of coil performance and eddy current effects. It was found that a flanged-edge coil design produced a better coil performance and more moderate eddy currents than those of the other designs. It is hoped that this study will help to inform the design of optimal gradient coils for split MRI systems with a large central gap.
Copyright © 2012 Elsevier Inc. All rights reserved.
Similar articles
-
Simulation and analysis of the interactions between split gradient coils and a split magnet cryostat in an MRI-PET system.J Magn Reson. 2012 Sep;222:8-15. doi: 10.1016/j.jmr.2012.05.022. Epub 2012 Jun 13. J Magn Reson. 2012. PMID: 22820005
-
Mitigation of Intra-coil Eddy Currents in Split Gradient Coils in a Hybrid MRI-LINAC System.IEEE Trans Biomed Eng. 2017 Mar;64(3):725-732. doi: 10.1109/TBME.2016.2573316. Epub 2016 May 26. IEEE Trans Biomed Eng. 2017. PMID: 27249823
-
Intra-coil interactions in split gradient coils in a hybrid MRI-LINAC system.J Magn Reson. 2016 Apr;265:52-8. doi: 10.1016/j.jmr.2016.01.013. Epub 2016 Jan 28. J Magn Reson. 2016. PMID: 26852418
-
Unconventional gradient coil designs in magnetic resonance imaging.Crit Rev Biomed Eng. 2014;42(6):493-526. doi: 10.1615/critrevbiomedeng.2015013100. Crit Rev Biomed Eng. 2014. PMID: 25955713 Review.
-
Engineering novel detectors and sensors for MRI.J Magn Reson. 2013 Apr;229:67-74. doi: 10.1016/j.jmr.2012.11.012. Epub 2012 Nov 29. J Magn Reson. 2013. PMID: 23245489 Free PMC article. Review.
Cited by
-
Gradient and shim technologies for ultra high field MRI.Neuroimage. 2018 Mar;168:59-70. doi: 10.1016/j.neuroimage.2016.11.033. Epub 2016 Nov 30. Neuroimage. 2018. PMID: 27915120 Free PMC article. Review.
-
Design and construction of gradient coils for an MRI-guided small animal radiation platform.Heliyon. 2024 Feb 19;10(4):e26251. doi: 10.1016/j.heliyon.2024.e26251. eCollection 2024 Feb 29. Heliyon. 2024. PMID: 39670066 Free PMC article.
-
Stereotactic radiosurgery planning of vestibular schwannomas: Is MRI at 3 Tesla geometrically accurate?Med Phys. 2017 Feb;44(2):375-381. doi: 10.1002/mp.12068. Med Phys. 2017. PMID: 28019663 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous