Validation of a stand-alone near-infrared spectroscopy system for monitoring cerebral autoregulation during cardiac surgery
- PMID: 23223100
- PMCID: PMC3800185
- DOI: 10.1213/ANE.0b013e318271fb10
Validation of a stand-alone near-infrared spectroscopy system for monitoring cerebral autoregulation during cardiac surgery
Abstract
Background: Individualizing arterial blood pressure (ABP) targets during cardiopulmonary bypass (CPB) based on cerebral blood flow (CBF) autoregulation monitoring may provide a more effective means for preventing cerebral hypoperfusion than the current standard of care. Autoregulation can be monitored in real time with transcranial Doppler (TCD). We have previously demonstrated that near-infrared spectroscopy (NIRS)-derived regional cerebral oxygen saturation (rS(c)O(2)) provides a clinically suitable surrogate of CBF for autoregulation monitoring. The purpose of this study was to determine the accuracy of a stand-alone "plug-and-play" investigational system for autoregulation monitoring that uses a commercially available NIRS monitor with TCD methods.
Methods: TCD monitoring of middle cerebral artery CBF velocity and NIRS monitoring were performed in 70 patients during CPB. Indices of autoregulation were computed by both a personal computer-based system and an investigational prototype NIRS-based monitor. A moving linear correlation coefficient between slow waves of ABP and CBF velocity (mean velocity index [Mx]) and between ABP and rS(c)O(2) (cerebral oximetry index [COx]) were calculated. When CBF is autoregulated, there is no correlation between CBF and ABP; when CBF is dysregulated, Mx and COx approach 1 (i.e., CBF and ABP are correlated). Linear regression and bias analysis were performed between time-averaged values of Mx and COx derived from the personal computer-based system and from COx measured with the prototype monitor. Values for Mx and COx were categorized in 5 mm Hg bins of ABP for each patient. The lower limit of CBF autoregulation was defined as the ABP where Mx incrementally increased to ≥0.4.
Results: There was correlation and good agreement between COx derived from the prototype monitor and Mx (r = 0.510; 95% confidence interval, 0.414-0.595; P < 0.001; bias, -0.07 ± 0.19). The correlation and bias between the personal computer-based COx and the COx from the prototype NIRS monitor were r = 0.957 (95% confidence interval, 0.945-0.966; P < 0.001 and 0.06 ± 0.06, respectively). The average ABP at the lower limit of autoregulation was 63 ± 11 mm Hg (95% prediction interval, 52-74 mm Hg). Although the mean ABP at the COx-determined lower limit of autoregulation determined with the prototype monitor was statistically different from that determined by Mx (59 ± 9 mm Hg; 95% prediction interval, 50-68 mm Hg; P = 0.026), the difference was not likely clinically meaningful.
Conclusions: Monitoring CBF autoregulation with an investigational stand-alone NIRS monitor is correlated and in good agreement with TCD-based methods. The availability of such a device would allow widespread autoregulation monitoring as a means of individualizing ABP targets during CPB.
Trial registration: ClinicalTrials.gov NCT00769691.
Figures



Similar articles
-
Validation of Near-Infrared Spectroscopy for Monitoring Cerebral Autoregulation in Comatose Patients.Neurocrit Care. 2017 Dec;27(3):362-369. doi: 10.1007/s12028-017-0421-8. Neurocrit Care. 2017. PMID: 28664392 Free PMC article.
-
Cerebral Autoregulation Monitoring with Ultrasound-Tagged Near-Infrared Spectroscopy in Cardiac Surgery Patients.Anesth Analg. 2015 Nov;121(5):1187-93. doi: 10.1213/ANE.0000000000000930. Anesth Analg. 2015. PMID: 26334746 Free PMC article. Clinical Trial.
-
Real-time continuous monitoring of cerebral blood flow autoregulation using near-infrared spectroscopy in patients undergoing cardiopulmonary bypass.Stroke. 2010 Sep;41(9):1951-6. doi: 10.1161/STROKEAHA.109.575159. Epub 2010 Jul 22. Stroke. 2010. PMID: 20651274 Free PMC article.
-
Noninvasive autoregulation monitoring with and without intracranial pressure in the naive piglet brain.Anesth Analg. 2010 Jul;111(1):191-5. doi: 10.1213/ANE.0b013e3181e054ba. Epub 2010 Jun 2. Anesth Analg. 2010. PMID: 20519421 Free PMC article. Review.
-
Monitoring of cerebral blood flow autoregulation: physiologic basis, measurement, and clinical implications.Br J Anaesth. 2024 Jun;132(6):1260-1273. doi: 10.1016/j.bja.2024.01.043. Epub 2024 Mar 12. Br J Anaesth. 2024. PMID: 38471987 Review.
Cited by
-
Perioperative organ injury.Anesthesiology. 2013 Dec;119(6):1474-89. doi: 10.1097/ALN.0000000000000022. Anesthesiology. 2013. PMID: 24126264 Free PMC article. Review.
-
Near-infrared Spectroscopy-derived Cerebral Autoregulation Indices Independently Predict Clinical Outcome in Acutely Ill Comatose Patients.J Neurosurg Anesthesiol. 2020 Jul;32(3):234-241. doi: 10.1097/ANA.0000000000000589. J Neurosurg Anesthesiol. 2020. PMID: 30864999 Free PMC article.
-
Dynamic cerebral autoregulation measured by diffuse correlation spectroscopy.J Cereb Blood Flow Metab. 2023 Aug;43(8):1317-1327. doi: 10.1177/0271678X231153728. Epub 2023 Jan 26. J Cereb Blood Flow Metab. 2023. PMID: 36703572 Free PMC article.
-
Monitoring cerebral ischemia using cerebral oximetry: pros and cons.J Biomed Res. 2016 Jan;30(1):1-4. doi: 10.7555/JBR.30.20150096. Epub 2015 Oct 28. J Biomed Res. 2016. PMID: 26679787 Free PMC article. No abstract available.
-
Cerebrovascular Autoregulation-Based Optimal Mean Arterial Pressure During Prostate Surgery - A Secondary Analysis of a Prospective Cohort Study.Ther Clin Risk Manag. 2025 May 27;21:757-767. doi: 10.2147/TCRM.S505676. eCollection 2025. Ther Clin Risk Manag. 2025. PMID: 40453324 Free PMC article.
References
-
- Czonsnyka M, Brady K, Reinhard M, Smielewski P, Steiner L. Monitoring of cerebrovascular autoregulation: Facts, Myths, and Missing Links. Neurocrit Care. 2009;10:373–86. - PubMed
-
- Reinhard M, Roth M. Effect of carotid endarterectomy or stenting on impairment of dynamic cerebral autoregulation. Stroke. 2004;35:1381–7. - PubMed
-
- Minhas PS, Smielewski P, Kirkpatrick PJ, Pickard JD, Czosnyka M. Pressure autoregulation and positron emission tomography-derived cerebral blood flow acetazolamide reactivity in patients with carotid artery stenosis. Neurosurg. 2004;55:63–8. - PubMed
Publication types
MeSH terms
Associated data
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical