High-pitch dual-source CT angiography of supra-aortic arteries: assessment of image quality and radiation dose
- PMID: 23223824
- DOI: 10.1007/s00234-012-1120-y
High-pitch dual-source CT angiography of supra-aortic arteries: assessment of image quality and radiation dose
Abstract
Introduction: High-pitch CT angiography (CTA) is a recent innovation that allows significant shortening of scan time with volume coverage of 43 mm per second. The aim of our study was to assess this technique in CTA of the head and neck.
Methods: CTA of supra-aortic arteries was performed in 50 patients using two acquisition protocols: conventional single-source 64-slice (pitch 1.2) and high-pitch dual-source 128-slice CT (pitch 3.2). Subjective and objective image quality of supra-aortic vessel ostia as well as intra- and extra-cranial segments was retrospectively assessed by blinded readers and radiation dose compared between the two protocols.
Results: Conventional and high-pitch CTA achieved comparable signal-to-noise ratios in arterial (54.3 ± 16.5 versus 57.3 ± 14.8; p = 0.50) and venous segments (15.8 ± 6.7 versus 18.9 ± 8.9; p = 0.21). High-pitch scanning was, however, associated with sharper delineation of vessel contours and image quality significantly improved at the level of supra-aortic vessel ostia (p < 0.0001) as well as along the brachiocephalic trunk (p < 0.0001), the subclavian arteries (p < 0.0001), proximal common carotid arteries (p = 0.01), and vertebral V1 segments (p < 0.0001). Using the high-pitch mode, the dose-length product was reduced by about 35% (218.2 ± 30 versus 141.8 ± 20 mGy × cm).
Conclusions: Due to elimination of transmitted cardiac motion, high-pitch CTA of the neck improves image quality in the proximity of the aortic arch while significantly lowering radiation dose. The technique thus qualifies as a promising alternative to conventional spiral CTA and may be particularly useful for identification of ostial stenosis.
Similar articles
-
Improved Pulmonary Artery Evaluation Using High-Pitch Photon-Counting CT Compared to High-Pitch Conventional or Routine-Pitch Conventional Dual-Energy CT.J Comput Assist Tomogr. 2024 Nov-Dec 01;48(6):897-905. doi: 10.1097/RCT.0000000000001645. Epub 2024 Aug 16. J Comput Assist Tomogr. 2024. PMID: 39146219
-
High-Pitch Dual-Source Computed Tomography Renal Angiography Comparison With Conventional Low-Pitch Computed Tomography Angiography: Image Quality, Contrast Medium Volume, and Radiation Dose.J Comput Assist Tomogr. 2015 Sep-Oct;39(5):737-40. doi: 10.1097/RCT.0000000000000268. J Comput Assist Tomogr. 2015. PMID: 26035407
-
Low-concentration contrast medium for 128-slice dual-source CT coronary angiography at a very low radiation dose using prospectively ECG-triggered high-pitch spiral acquisition.Acad Radiol. 2015 Feb;22(2):195-202. doi: 10.1016/j.acra.2014.07.025. Epub 2014 Nov 11. Acad Radiol. 2015. PMID: 25457735 Clinical Trial.
-
3D CT angiography.Crit Rev Diagn Imaging. 2001 Feb;42(1):1-28. Crit Rev Diagn Imaging. 2001. PMID: 11244908 Review. No abstract available.
-
Dual-source computed tomography protocols for the pediatric chest - scan optimization techniques.Pediatr Radiol. 2023 Jun;53(7):1248-1259. doi: 10.1007/s00247-022-05468-7. Epub 2022 Aug 11. Pediatr Radiol. 2023. PMID: 35948645 Free PMC article. Review.
Cited by
-
Assessment of carotid plaque composition using fast-kV switching dual-energy CT with gemstone detector: comparison with extracorporeal and virtual histology-intravascular ultrasound.Neuroradiology. 2015 Sep;57(9):889-95. doi: 10.1007/s00234-015-1541-5. Epub 2015 May 16. Neuroradiology. 2015. PMID: 25981057
-
Automated tube voltage adaptation in head and neck computed tomography between 120 and 100 kV: effects on image quality and radiation dose.Neuroradiology. 2014 Sep;56(9):797-803. doi: 10.1007/s00234-014-1393-4. Epub 2014 Jun 25. Neuroradiology. 2014. PMID: 24961883
-
Low-dose high-pitch CT angiography of the supraaortic arteries using sinogram-affirmed iterative reconstruction.PLoS One. 2014 Jun 11;9(6):e99832. doi: 10.1371/journal.pone.0099832. eCollection 2014. PLoS One. 2014. PMID: 24919195 Free PMC article. Clinical Trial.
-
Diagnostic accuracy of coronary CT angiography combined with dual-energy myocardial perfusion imaging for detection of myocardial infarction.Exp Ther Med. 2017 Jul;14(1):207-213. doi: 10.3892/etm.2017.4485. Epub 2017 May 22. Exp Ther Med. 2017. PMID: 28672916 Free PMC article.
-
Variation of degree of stenosis quantification using different energy level with dual energy CT scanner.Neuroradiology. 2019 Mar;61(3):285-291. doi: 10.1007/s00234-018-2142-x. Epub 2018 Dec 15. Neuroradiology. 2019. PMID: 30554271