Model-based iterative reconstruction in pediatric chest CT: assessment of image quality in a prospective study of children with cystic fibrosis

Affiliations

Model-based iterative reconstruction in pediatric chest CT: assessment of image quality in a prospective study of children with cystic fibrosis

Frédéric A Miéville et al. Pediatr Radiol. 2013 Mar.

Abstract

Background: The potential effects of ionizing radiation are of particular concern in children. The model-based iterative reconstruction VEO(TM) is a technique commercialized to improve image quality and reduce noise compared with the filtered back-projection (FBP) method.

Objective: To evaluate the potential of VEO(TM) on diagnostic image quality and dose reduction in pediatric chest CT examinations.

Materials and methods: Twenty children (mean 11.4 years) with cystic fibrosis underwent either a standard CT or a moderately reduced-dose CT plus a minimum-dose CT performed at 100 kVp. Reduced-dose CT examinations consisted of two consecutive acquisitions: one moderately reduced-dose CT with increased noise index (NI = 70) and one minimum-dose CT at CTDIvol 0.14 mGy. Standard CTs were reconstructed using the FBP method while low-dose CTs were reconstructed using FBP and VEO. Two senior radiologists evaluated diagnostic image quality independently by scoring anatomical structures using a four-point scale (1 = excellent, 2 = clear, 3 = diminished, 4 = non-diagnostic). Standard deviation (SD) and signal-to-noise ratio (SNR) were also computed.

Results: At moderately reduced doses, VEO images had significantly lower SD (P < 0.001) and higher SNR (P < 0.05) in comparison to filtered back-projection images. Further improvements were obtained at minimum-dose CT. The best diagnostic image quality was obtained with VEO at minimum-dose CT for the small structures (subpleural vessels and lung fissures) (P < 0.001). The potential for dose reduction was dependent on the diagnostic task because of the modification of the image texture produced by this reconstruction.

Conclusions: At minimum-dose CT, VEO enables important dose reduction depending on the clinical indication and makes visible certain small structures that were not perceptible with filtered back-projection.

PubMed Disclaimer

Similar articles

Cited by

References

    1. AJR Am J Roentgenol. 2010 Nov;195(5):1095-9 - PubMed
    1. Phys Med. 2013 Jan;29(1):99-110 - PubMed
    1. AJR Am J Roentgenol. 2010 Jul;195(1):126-31 - PubMed
    1. Eur Radiol. 2011 Mar;21(3):636-43 - PubMed
    1. Radiology. 2010 Jan;254(1):145-53 - PubMed

Publication types