Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2013 Mar;21(3):127-34.
doi: 10.1007/s12471-012-0341-3.

Long-term performance of the St Jude Riata 1580-1582 ICD lead family

Affiliations

Long-term performance of the St Jude Riata 1580-1582 ICD lead family

S D A Valk et al. Neth Heart J. 2013 Mar.

Abstract

Objective: Safety concerns about the Riata ICD shock lead were recently raised, with insulation failure due to conductor externalisation. Its incidence and presentation were assessed, and predictors of insulation failure and lead survival of the Riata 1580-1582 were studied, retrospectively, before the official recall.

Methods: All 374 patients at the Erasmus Medical Center between July 2003 and December 2007 with a 1580, 1581 or 1582 shock lead.

Results: The majority of the patients were male (78 %), with a median age of 60 years (IQR 52-70); primary prevention in 61 %. Median follow-up was 60.3 months (IQR 35.5-73.2), with 117 (31 %) patients dying. Electrical abnormalities (mainly noise, 65 %) were observed in 20/257 patients (7.8 %). Definite conductor externalisation was confirmed with fluoroscopy or chest X-ray in 16 patients, and in one after extraction. One patient presented with a drop in the high-voltage impedance trend with a short circuit of the ICD system during defibrillation testing, and needed to be shocked externally. In 8 more patients, conductor externalisation was found during an elective procedure. No predictors of externalisation could be found, except for the use of single coil (p = 0.02). Median time to conductor externalisation was 5 years (IQR 3.1-6.2). Lead externalisation was observed in 5.4 % (95 % CI 3.1-9.3) at 5 years and 22.7 % (95 % CI 13.6-36.6) at 8 years.

Conclusion: A high incidence of insulation defects associated with conductor externalisation in the Riata ICD lead family is observed. The mode of presentation is diverse. This type of insulation failure can lead to failure of therapy delivery.

PubMed Disclaimer

Figures

Fig. 1
Fig. 1
At the left: lateral view; fluoroscopic image of an externalised conductor in patient 9 (which led to a defect ICD after a high-voltage shock). At the right a lateral view with enlargement of a lead segment which is near normal. This patient had an electrical defect with noise, and could no longer be paced
Fig. 2
Fig. 2
Segment of the lead showing 4 externalised conductors which had abrased the outer isolation
Fig. 3
Fig. 3
Proportion of externalised conductors as appearing over time with 95 % confidence intervals

References

    1. Epstein AE, DiMarco JP, Ellenbogen KA, et al. American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices); American Association for Thoracic Surgery; Society of Thoracic Surgeons. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices): developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons. Circulation. 2008;117(21):e350–e408. doi: 10.1161/CIRCUALTIONAHA.108.189742. - DOI - PubMed
    1. Cannom DS, Prystowsky EN. The evolution of the implantable cardioverter defibrillator. NASPE history series. Pacing Clin Electropysiol. 2004;27:419–431. doi: 10.1111/j.1540-8159.2004.00457.x. - DOI - PubMed
    1. Camm AJ, Nisam S. European utilization of the implantable defibrillator: has 10 years changed the ‘enigma’? Europace. 2010;12:1063–1069. doi: 10.1093/europace/euq282. - DOI - PubMed
    1. Hauser RG. The growing mismatch between patient longevity and the service life of implantable cardioverter-defibrillators. J Am Coll Cardiol. 2005;45:2022–2025. doi: 10.1016/j.jacc.2005.02.077. - DOI - PubMed
    1. Kleemann T, Becker T, Doenges K, et al. Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of >10 years. Circulation. 2007;115:2474–2480. doi: 10.1161/CIRCULATIONAHA.106.663807. - DOI - PubMed

LinkOut - more resources