Long-term performance of the St Jude Riata 1580-1582 ICD lead family
- PMID: 23229809
- PMCID: PMC3578533
- DOI: 10.1007/s12471-012-0341-3
Long-term performance of the St Jude Riata 1580-1582 ICD lead family
Abstract
Objective: Safety concerns about the Riata ICD shock lead were recently raised, with insulation failure due to conductor externalisation. Its incidence and presentation were assessed, and predictors of insulation failure and lead survival of the Riata 1580-1582 were studied, retrospectively, before the official recall.
Methods: All 374 patients at the Erasmus Medical Center between July 2003 and December 2007 with a 1580, 1581 or 1582 shock lead.
Results: The majority of the patients were male (78 %), with a median age of 60 years (IQR 52-70); primary prevention in 61 %. Median follow-up was 60.3 months (IQR 35.5-73.2), with 117 (31 %) patients dying. Electrical abnormalities (mainly noise, 65 %) were observed in 20/257 patients (7.8 %). Definite conductor externalisation was confirmed with fluoroscopy or chest X-ray in 16 patients, and in one after extraction. One patient presented with a drop in the high-voltage impedance trend with a short circuit of the ICD system during defibrillation testing, and needed to be shocked externally. In 8 more patients, conductor externalisation was found during an elective procedure. No predictors of externalisation could be found, except for the use of single coil (p = 0.02). Median time to conductor externalisation was 5 years (IQR 3.1-6.2). Lead externalisation was observed in 5.4 % (95 % CI 3.1-9.3) at 5 years and 22.7 % (95 % CI 13.6-36.6) at 8 years.
Conclusion: A high incidence of insulation defects associated with conductor externalisation in the Riata ICD lead family is observed. The mode of presentation is diverse. This type of insulation failure can lead to failure of therapy delivery.
Figures
References
-
- Epstein AE, DiMarco JP, Ellenbogen KA, et al. American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices); American Association for Thoracic Surgery; Society of Thoracic Surgeons. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices): developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons. Circulation. 2008;117(21):e350–e408. doi: 10.1161/CIRCUALTIONAHA.108.189742. - DOI - PubMed
LinkOut - more resources
Full Text Sources