Influence of scan mode (180°/360°) of the cone beam computed tomography for preoperative dental implant measurements
- PMID: 23231370
- DOI: 10.1111/clr.12080
Influence of scan mode (180°/360°) of the cone beam computed tomography for preoperative dental implant measurements
Abstract
Objective: The aim of this study was to evaluate the effect of scan mode of the cone beam computed tomography (CBCT) in the preoperative dental implant measurements.
Material and methods: Completely edentulous mandibles with entirely resorbed alveolar processes were selected for this study. Five regions were selected (incisor, canine, premolar, first molar, and second molar). The mandibles were scanned with Next Generation i-CAT CBCT unit (Imaging Sciences International, Inc, Hatfield, PA, USA) with half (180°) and full (360°) mode. Two oral radiologists performed vertical measurements in all selected regions; the measurements of half of the sample were repeated within an interval of 30 days. The mandibles were sectioned using an electrical saw in all evaluated regions to obtain the gold standard. The intraclass correlation coefficient was calculated for the intra- and interobserver agreement. Descriptive statistics were calculated as mean, median, and standard deviation. Wilcoxon signed rank test was used to determine the correlation between the measurements obtained in different scan mode with the gold standard. The significance level was 5%.
Results: The values of intra- and interobserver reproducibility indicated a strong agreement. In the dental implant measurements, except the bone height of the second molar region in full scan mode (P = 0.02), the Wilcoxon signed rank test did not show statistical significant difference with the gold standard (P > 0.05).
Conclusions: Both modes provided real measures, necessary when performing implant planning; however, half scan mode uses smaller doses, following the principle of effectiveness. We believe that this method should be used because of the best dose-effect relationship and offer less risk to the patient.
Keywords: cone beam computed tomography; dental implants; mandible; measurements.
© 2012 John Wiley & Sons A/S.
Similar articles
-
Accuracy of linear measurements in cone beam computed tomography with different voxel sizes.Implant Dent. 2012 Apr;21(2):150-5. doi: 10.1097/ID.0b013e31824bf93c. Implant Dent. 2012. PMID: 22382754
-
Vertical bone measurements from cone beam computed tomography images using different software packages.Braz Oral Res. 2015;29:S1806-83242015000100236. doi: 10.1590/1807-3107BOR-2015.vol29.0035. Epub 2015 Feb 11. Braz Oral Res. 2015. PMID: 25715034
-
Influence of the milliamperage settings on cone beam computed tomography imaging for implant planning.Int J Oral Maxillofac Implants. 2014 Nov-Dec;29(6):1364-8. doi: 10.11607/jomi.3524. Epub 2014 Sep 26. Int J Oral Maxillofac Implants. 2014. PMID: 25265127
-
The incisive canal. Considerations during implant placement: case report and literature review.Int J Oral Maxillofac Implants. 2009 Jul-Aug;24(4):740-5. Int J Oral Maxillofac Implants. 2009. PMID: 19885417 Review.
-
Accuracy of linear measurements on CBCT images related to presurgical implant treatment planning: A systematic review.Clin Oral Implants Res. 2018 Oct;29 Suppl 16:393-415. doi: 10.1111/clr.13142. Clin Oral Implants Res. 2018. PMID: 30328204
Cited by
-
Etiology and Measurement of Peri-Implant Crestal Bone Loss (CBL).J Clin Med. 2019 Feb 1;8(2):166. doi: 10.3390/jcm8020166. J Clin Med. 2019. PMID: 30717189 Free PMC article. Review.
-
Effect of Dental Implant Metal Artifacts on Accuracy of Linear Measurements by Two Cone-Beam Computed Tomography Systems Before and After Crown Restoration.J Dent (Tehran). 2017 Nov;14(6):329-336. J Dent (Tehran). 2017. PMID: 29942327 Free PMC article.
-
Feasibility of micro-computed tomography to detect and classify proximal caries lesions in vitro.Dent Res J (Isfahan). 2018 Mar-Apr;15(2):123-129. Dent Res J (Isfahan). 2018. PMID: 29576776 Free PMC article.
-
Accuracy of linear measurements for implant planning based on low-dose cone beam CT protocols: a systematic review and meta-analysis.Dentomaxillofac Radiol. 2024 Apr 29;53(4):207-221. doi: 10.1093/dmfr/twae007. Dentomaxillofac Radiol. 2024. PMID: 38429951 Free PMC article.
-
Novel low-dose protocols using cone beam computed tomography in dental medicine: a review focusing on indications, limitations, and future possibilities.Clin Oral Investig. 2019 Jun;23(6):2573-2581. doi: 10.1007/s00784-019-02907-y. Epub 2019 Apr 25. Clin Oral Investig. 2019. PMID: 31025192 Review.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous