Free-breathing 3 T magnetic resonance T2-mapping of the heart

Affiliations
Free article
Comparative Study

Free-breathing 3 T magnetic resonance T2-mapping of the heart

Ruud B van Heeswijk et al. JACC Cardiovasc Imaging. 2012 Dec.
Free article

Abstract

Objectives: This study sought to establish an accurate and reproducible T(2)-mapping cardiac magnetic resonance (CMR) methodology at 3 T and to evaluate it in healthy volunteers and patients with myocardial infarct.

Background: Myocardial edema affects the T(2) relaxation time on CMR. Therefore, T(2)-mapping has been established to characterize edema at 1.5 T. A 3 T implementation designed for longitudinal studies and aimed at guiding and monitoring therapy remains to be implemented, thoroughly characterized, and evaluated in vivo.

Methods: A free-breathing navigator-gated radial CMR pulse sequence with an adiabatic T(2) preparation module and an empirical fitting equation for T(2) quantification was optimized using numerical simulations and was validated at 3 T in a phantom study. Its reproducibility for myocardial T(2) quantification was then ascertained in healthy volunteers and improved using an external reference phantom with known T(2). In a small cohort of patients with established myocardial infarction, the local T(2) value and extent of the edematous region were determined and compared with conventional T(2)-weighted CMR and x-ray coronary angiography, where available.

Results: The numerical simulations and phantom study demonstrated that the empirical fitting equation is significantly more accurate for T(2) quantification than that for the more conventional exponential decay. The volunteer study consistently demonstrated a reproducibility error as low as 2 ± 1% using the external reference phantom and an average myocardial T(2) of 38.5 ± 4.5 ms. Intraobserver and interobserver variability in the volunteers were -0.04 ± 0.89 ms (p = 0.86) and -0.23 ± 0.91 ms (p = 0.87), respectively. In the infarction patients, the T(2) in edema was 62.4 ± 9.2 ms and was consistent with the x-ray angiographic findings. Simultaneously, the extent of the edematous region by T(2)-mapping correlated well with that from the T(2)-weighted images (r = 0.91).

Conclusions: The new, well-characterized 3 T methodology enables robust and accurate cardiac T(2)-mapping at 3 T with high spatial resolution, while the addition of a reference phantom improves reproducibility. This technique may be well suited for longitudinal studies in patients with suspected or established heart disease.