Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2013 Apr;131(4):480e-489e.
doi: 10.1097/PRS.0b013e3182818a00.

Rupture of poly implant prothèse silicone breast implants: an implant retrieval study

Affiliations

Rupture of poly implant prothèse silicone breast implants: an implant retrieval study

Eric Swarts et al. Plast Reconstr Surg. 2013 Apr.

Abstract

Background: Poly Implant Prothèse implants were recalled in Australia in April of 2010 following concerns of higher than expected rupture rates and the use of unauthorized industrial grade silicone as a filler material. Although subsequent investigations found that the gel filler material does not pose a threat to human health, the important question of what caused a relatively modern breast implant to have such a poor outcome compared with contemporary silicone breast implants is yet to be addressed.

Methods: From a cohort of 27 patients, 19 ruptured Poly Implant Prothèse breast implants were subjected to a range of mechanical tests and microscopic/macroscopic investigations to evaluate possible changes in properties as a result of implantation. New Poly Implant Prothèse implants were used as controls.

Results: All samples, explanted and controls, complied with the requirements for shell integrity as specified in the International Organization for Standardization 14607. Compression testing revealed rupture rates similar to those reported in the literature. Shell thickness was highly variable, with most shells having regions below the minimum thickness of 0.57 mm that was specified by the manufacturer. Potential regions of stress concentration were observed on the smooth inner surfaces and outer textured surfaces.

Conclusions: The high incidence of Poly Implant Prothèse shell rupture is most likely a result of inadequate quality control, with contributory factors being shell thickness variation and manufacturing defects on both inner and outer surfaces of the shell. No evidence of shell degradation with implantation time was determined.

PubMed Disclaimer

References

    1. Brandon HJ, Jerina KL, Wolf CJ, Young VL. Biodurability of retrieved silicone gel breast implants. Plast Reconstr Surg. 2003;111:2295–2306.
    1. Brandon HJ, Young VL, Watson ME, Wolf CJ, Jerina KL. Protocol for retrieval and analysis of breast implants. J Long Term Eff Med Implants 2003;13:49–61.
    1. Berry MG, Stanek JJ. The PIP mammary prosthesis: A product recall study. J Plast Reconstr Aesthet Surg. 2012;65:697–704.
    1. Maijers MC, Niessen FB. Prevalence of rupture in Poly Implant Prothèse silicone breast implants, recalled from the European market in 2010. Plast Reconstr Surg. 2012;129:1372–1378.
    1. Barr S, Hill E, Bayat A. Current implant surface technology: An examination of their nanostructure and their influence on fibroblast alignment and biocompatibility. Eplasty 2009;9:198–217.

LinkOut - more resources