Stability of AMH measurement in blood and avoidance of proteolytic changes
- PMID: 23260035
- DOI: 10.1016/j.rbmo.2012.11.005
Stability of AMH measurement in blood and avoidance of proteolytic changes
Abstract
The new Gen II assay for anti-Müllerian hormone (AMH) shows good stability and reliability in serum, but analyses of stability in whole blood are lacking. Testing the effects of storage of whole-blood samples at room temperature revealed significant increases in the measured value of AMH of 31% over 4 days (P<0.001). The effect is temperature dependent, with storage at 4°C showing markedly reduced increments. Further, samples collected into serum tubes with gel separators and centrifuged within 5h (blood cells and serum physically separated within the collection tube) showed reliable stability over a period of more than 5 days.
Copyright © 2012 Reproductive Healthcare Ltd. Published by Elsevier Ltd. All rights reserved.