Review
. 2013 Summer;38(2):133-42.
doi: 10.1016/j.meddos.2012.10.001. Epub 2012 Dec 21.

Retrospective review of Contura HDR breast cases to improve our standardized procedure

Affiliations
Review

Retrospective review of Contura HDR breast cases to improve our standardized procedure

Ileana Iftimia et al. Med Dosim. 2013 Summer.

Abstract

To retrospectively review our first 20 Contura high dose rate breast cases to improve and refine our standardized procedure and checklists. We prepared in advance checklists for all steps, developed an in-house Excel spreadsheet for second checking the plan, and generated a procedure for efficient contouring and a set of optimization constraints to meet the dose volume histogram criteria. Templates were created in our treatment planning system for structures, isodose levels, optimization constraints, and plan report. This study reviews our first 20 high dose rate Contura breast treatment plans. We followed our standardized procedure for contouring, planning, and second checking. The established dose volume histogram criteria were successfully met for all plans. For the cases studied here, the balloon-skin and balloon-ribs distances ranged between 5 and 43 mm and 1 and 33 mm, respectively; air_seroma volume/PTV_Eval volume≤5.5% (allowed≤10%); asymmetry<1.2mm (goal≤2 mm); PTV_Eval V90%≥97.6%; PTV_Eval V95%≥94.9%; skin max dose≤98%Rx; ribs max dose≤137%Rx; V150%≤29.8 cc; V200%≤7.8 cc; the total dwell time range was 225.4 to 401.9 seconds; and the second check agreement was within 3%. Based on this analysis, more appropriate ranges for the total dwell time and balloon diameter tolerance were found. Three major problems were encountered: balloon migration toward the skin for small balloon-to-skin distances, lumen obstruction, and length change for the flexible balloon. Solutions were found for these issues and our standardized procedure and checklists were updated accordingly. Based on our review of these cases, the use of checklists resulted in consistent results, indicating good coverage for the target without sacrificing the critical structures. This review helped us to refine our standardized procedure and update our checklists.