Implantation of personalized, biocompatible mitral annuloplasty rings: feasibility study in an animal model
- PMID: 23287589
- PMCID: PMC3598035
- DOI: 10.1093/icvts/ivs531
Implantation of personalized, biocompatible mitral annuloplasty rings: feasibility study in an animal model
Abstract
Objectives: Implantation of an annuloplasty ring is an essential component of a durable mitral valve repair. Currently available off-the-shelf rings still do not cover all the variations in mitral annulus anatomy and pathology from subject to subject. Computed tomography (CT) and echo imaging allow for 3-D segmentation of the mitral valve and mitral annulus. The concept of tailored annuloplasty rings has been proposed although, to date, no surgically applicable implementation of patient-specific annuloplasty rings has been seen. The objective of this trial was to prove the concept of surgical implantation of a model-guided, personalized mitral annuloplasty ring, manufactured based on individual CT-scan models.
Methods: ECG-gated CT angiography was performed in six healthy pigs under general anaesthesia. Based on the individual shape of the mitral annulus in systole, a customized solid ring with integrated suturing holes was designed and manufactured from a biocompatible titanium alloy by a rapid process using laser melting. The ring was implanted three days later and valve function was assessed by intraoperative echocardiography. The macroscopic annulus-annuloplasty match was assessed after heart explantation.
Results: CT angiography provided good enough image quality in all animals to allow for segmentation of the mitral annulus. The individually tailored mitral rings were manufactured and successfully implanted in all pigs. In 50%, a perfect matching of the implanted ring and the mitral annulus was achieved. In one animal, a slight deviation of the ring shape from the circumference was seen postoperatively. The rings implanted in the first two animals were significantly oversized but the deviation did not affect valve competence.
Conclusions: CT image quality and accuracy of the dimensions of the mitral annulus were sufficient for digital modelling and rapid manufacturing of mitral rings. Implantation of individually tailored annuloplasty rings is feasible.
Figures



Comment in
-
eComment. Personalized annuloplasty rings for mitral valve repair: lessons from the pig model.Interact Cardiovasc Thorac Surg. 2013 Apr;16(4):422. doi: 10.1093/icvts/ivt069. Interact Cardiovasc Thorac Surg. 2013. PMID: 23504747 Free PMC article. No abstract available.
Similar articles
-
The geometrical effect of different annuloplasty rings on mitral valve annulus.J Cardiovasc Surg (Torino). 2017 Jun;58(3):481-488. doi: 10.23736/S0021-9509.16.09314-9. Epub 2016 May 6. J Cardiovasc Surg (Torino). 2017. PMID: 27153370
-
Do annuloplasty rings designed to treat ischemic/functional mitral regurgitation alter left-ventricular dimensions in the acutely ischemic ovine heart?J Thorac Cardiovasc Surg. 2019 Oct;158(4):1058-1068. doi: 10.1016/j.jtcvs.2018.12.077. Epub 2019 Jan 11. J Thorac Cardiovasc Surg. 2019. PMID: 30803776
-
Remodeling Mitral Annuloplasty Ring Concept with Preserved Dynamics of Annular Height.J Heart Valve Dis. 2017 May;26(3):295-303. J Heart Valve Dis. 2017. PMID: 29092114
-
Transcatheter Mitral Valve Direct Annuloplasty with the Millipede IRIS Ring.Interv Cardiol Clin. 2019 Jul;8(3):261-267. doi: 10.1016/j.iccl.2019.02.001. Epub 2019 Apr 8. Interv Cardiol Clin. 2019. PMID: 31078181 Review.
-
Systematic evaluation of the flexible and rigid annuloplasty ring after mitral valve repair for mitral regurgitation.Eur J Cardiothorac Surg. 2011 Aug;40(2):480-7. doi: 10.1016/j.ejcts.2010.11.080. Epub 2011 Feb 3. Eur J Cardiothorac Surg. 2011. PMID: 21295489
Cited by
-
eComment. Personalized annuloplasty rings for mitral valve repair: lessons from the pig model.Interact Cardiovasc Thorac Surg. 2013 Apr;16(4):422. doi: 10.1093/icvts/ivt069. Interact Cardiovasc Thorac Surg. 2013. PMID: 23504747 Free PMC article. No abstract available.
-
Stereolithography for Personalized Left Atrial Appendage Occluders.Adv Mater Technol. 2018 Dec;3(12):1800233. doi: 10.1002/admt.201800233. Epub 2018 Sep 21. Adv Mater Technol. 2018. PMID: 31803824 Free PMC article.
-
Using a shape prior for robust modeling of the mitral annulus on 4D ultrasound data.Int J Comput Assist Radiol Surg. 2014 Jul;9(4):635-44. doi: 10.1007/s11548-013-0942-3. Int J Comput Assist Radiol Surg. 2014. PMID: 24122458
-
Computed tomography for planning and postoperative imaging of transvenous mitral annuloplasty: first experience in an animal model.Int J Cardiovasc Imaging. 2015 Jan;31(1):135-42. doi: 10.1007/s10554-014-0516-7. Epub 2014 Aug 15. Int J Cardiovasc Imaging. 2015. PMID: 25119889
-
Sizing the mitral annulus in healthy subjects and patients with mitral regurgitation: 2D versus 3D measurements from cardiac CT.Int J Cardiovasc Imaging. 2014 Feb;30(2):389-98. doi: 10.1007/s10554-013-0341-4. Epub 2013 Dec 4. Int J Cardiovasc Imaging. 2014. PMID: 24306054
References
-
- Vahanian A, Alfieri O, Andreotti F, Antunes MJ, Barón-Esquivias G, et al. Authors/Task Force Members. Guidelines on the management of valvular heart disease (version 2012): The Joint Task Force on the Management of Valvular Heart Disease of the European Society of Cardiology (ESC) and the European Association for Cardio-Thoracic Surgery (EACTS) Eur J Cardiothorac Surg. 2012;42:S1–44. - PubMed
-
- Markar SR, Sadat U, Edmonds L, Nair SK. Mitral valve repair versus replacement in the elderly population. J Heart Valve Dis. 2011;20:265–71. - PubMed
-
- Seeburger J, Borger MA, Doll N, Walther T, Passage J, Falk V, et al. Comparison of outcomes of minimally invasive mitral valve surgery for posterior, anterior and bileaflet prolapse. Eur J Cardiothorac Surg. 2009;36:532–38. - PubMed
-
- David TE, Ivanov J, Armstrong S, Christie D, Rakowski H. A comparison of outcomes of mitral valve repair for degenerative disease with posterior, anterior, and bileaflet prolapse. J Thorac and Cardiovasc Surg. 2005;130:1242–49. - PubMed
-
- Braunberger E, Deloche A, Berrebi A, Abdallah F, Celestin JA, Meimoun P, et al. Very long-term results (more than 20 years) of valve repair with Carpentier's techniques in nonrheumatic mitral valve insufficiency. Circulation. 2001;104:I8–I11. - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources