Comparative Study
. 2013 Jan 7:12:6.
doi: 10.1186/1475-2875-12-6.

Diagnostic performance of CareStart™ malaria HRP2/pLDH (Pf/pan) combo test versus standard microscopy on falciparum and vivax malaria between China-Myanmar endemic borders

Affiliations
Comparative Study

Diagnostic performance of CareStart™ malaria HRP2/pLDH (Pf/pan) combo test versus standard microscopy on falciparum and vivax malaria between China-Myanmar endemic borders

Sun Xiaodong et al. Malar J. .

Abstract

Background: Rapid diagnostic test (RDT) is becoming an alternative way of establishing quickly the diagnosis of malaria infections, by detecting specific malaria antigens in suspected patients' blood between the China-Myanmar endemic borders areas, towards achieving the National Malaria Elimination programme by 2020. The objective of this study is to evaluate the performance of CareStart™ Malaria Pf/Pan RDT kit for the diagnosis of malaria infections in suspected patients. Blood examination by microscopy was taken as gold standard to evaluate CareStart™ kit's sensitivity, specificity and predictive value and corrected with PCR assay.

Results: Overall 126 of 241 (52.28%) malaria cases were detected by microscopy compared to 115 of 241(47.72%) CareStart™ kit and 128 of 241 (53.11%) PCR corrected assay. CareStart™ kit's sensitivity and specificity for the diagnosis of malaria were 89.68% and 98.26% respectively, compared to standard microscopy, whereas the sensitivity and specificity for falciparum malaria were 88.52% and 98.26%, and for vivax malaria: 90.77% and 100%. The CareStart™ positive predictive values were 98.26% (93.88-99.52%, 95% CI) compared to 100% (96.77-100%, 95% CI) for PCR-corrected, and the negative predictive values of 89.68% (83.15-93.87%, 95% CI) were the same in microscopy as PCR-corrected. The diagnostic accuracy of CareStart™ kit versus microscopy and PCR were 93.78% (89.99-96.19%, 95% CI) and 94.61% (90.99-96.82%, 95% CI) respectively. The likelihood of diagnostic of malaria positive was almost similar between microscopy and CareStart™ kit, with an entropy reduction of 60.0% compared to a weak likelihood of misdiagnosis of 0.10 (0.09-0.12, 95% CI), with an entropy reduction of 36.01%.

Conclusion: The accuracy of CareStart™ kit is comparable to gold standard microscopy in these areas, it is easy to perform and suitable for cross-border diagnosis and monitoring of local or imported malaria patterns by any local health staff in endemic remotes.

PubMed Disclaimer

References

    1. World Health Organization. Malaria rapid diagnostic test performance. Results of WHO product testing of malaria RDTs. Geneva: World Health Organisation; 2010.
    1. McMorrow ML, Masanja MI, Abdulla SM, Kahigwa E, Kachur SP. Challenges in routine implementation and quality control of rapid diagnostic tests for malaria-Rufiji District, Tanzania. Am J Trop Med Hyg. 2008;79:385–390. - PMC - PubMed
    1. Ministry of Health. Annual malaria reports. Beijing, People’s Republic of China; 2010.
    1. World Health Organization. Malaria rapid diagnostic test performance. Results of WHO product testing of malaria RDTs: round 1. Geneva: World Health Organization; 2009.
    1. Moody A. Rapid diagnostic tests for malaria parasites. Clin Microbiol Rev. 2002;15:66–78. doi: 10.1128/CMR.15.1.66-78.2002. - DOI - PMC - PubMed